2.4 Connecting the detector inlet capillary
■■■■■■■■■■■■■■■■■■■■■■
12
■■■■■■■■
2.850.9010 IC Conductivity Detector
Connecting detector inlet capillary to MSM / MSM
‑
HC
1 Connecting detector input
■
Connect the detector inlet capillary
and the capillary of the
MSM / MSM
‑
labeled with
out
to one another using a
6.2744.040 coupling
and two 6.2744.070 short pressure
screws
1
2
3
4
Figure 5
Connector detector – MSM
1
Detector inlet capillary
2
MSM / MSM
‑
HC outlet capillary
labeled with
out
.
3
Coupling 6.2744.040
4
Pressure screws, short 6.2744.070
Connecting the detector inlet capillary to the MCS
1 Connecting detector input
■
Fasten the detector inlet capillary
PEEK pressure screw
.
3
2
1
Figure 6
Connection detector – MCS
1
Detector inlet capillary
2
Pressure screw, long 6.2744.090
3
MCS output