IC Professional Detector
IC Conductivity Detector
Manual
8.850.8057EN / 2019-09-23
Page 1: ...IC Professional Detector IC Conductivity Detector Manual 8 850 8057EN 2019 09 23...
Page 2: ......
Page 3: ...rohm AG CH 9100 Herisau Switzerland Phone 41 71 353 85 85 Fax 41 71 353 89 01 info metrohm com www metrohm com IC Professional Detector IC Conductivity Detector 2 850 9010 Manual 8 850 8057EN 2019 09...
Page 4: ...echcom metrohm com This documentation is protected by copyright All rights reserved This documentation has been prepared with great care However errors can never be entirely ruled out Please send comm...
Page 5: ...iquids 6 1 5 4 Recycling and disposal 6 2 Installation 7 2 1 Setting up the instrument 7 2 1 1 Packaging 7 2 1 2 Check 7 2 1 3 Location 7 2 2 Inserting the detector 7 2 3 Connecting the detector 10 2...
Page 6: ...Table of contents IV 2 850 9010 IC Conductivity Detector 7 Accessories 19 Index 20...
Page 7: ...C Conductivity Detector V Table of figures Figure 1 Front 2 Figure 2 Rear 3 Figure 3 Inserting the detector 8 Figure 4 Connection detector separation column 11 Figure 5 Connector detector MSM 12 Figur...
Page 8: ......
Page 9: ...indicates these signals in digital form DSP Digital Signal Processing The conductivity detector exhibits outstanding thermal stability and thus guarantees reproducible measuring conditions The IC Con...
Page 10: ...instrument 2 2 850 9010 IC Conductivity Detector 1 2 Overview of the instrument 1 2 1 Front 2 1 3 Figure 1 Front 1 IC detector 2 850 9010 2 Opening for temperature sensor 3 Detector inlet capillary Pe...
Page 11: ...precise measurement of the conductivity during the ion chromatographic determination of cations and anions As the corresponding IC instrument the IC Conductivity Detector is also suitable for process...
Page 12: ...figure legend The first number refers to the figure number the sec ond to the instrument part in the figure Instruction step Carry out these steps in the sequence shown Warning This symbol draws atte...
Page 13: ...RNING Only personnel qualified by Metrohm are authorized to carry out service work on electronic components WARNING Never open the housing of the instrument The instrument could be damaged by this The...
Page 14: ...fluids and their disposal 1 5 4 Recycling and disposal This product is covered by European Directive 2012 19 EU WEEE Waste Electrical and Electronic Equipment The correct disposal of your old instrume...
Page 15: ...to ensure that it is complete and undamaged 2 1 3 Location The IC Conductivity Detector is designed for usage in the detector cham ber of the IC instruments For the location the same conditions apply...
Page 16: ...2 2 Inserting the detector 8 2 850 9010 IC Conductivity Detector 2 3 4 5 1 1 2 Figure 3 Inserting the detector 1 Detector connection cable 2 Detector outlet capillary...
Page 17: ...he detector Put the detector through the opening and position it on the sup port surface intended for this purpose and slide it as far as the limit stop to the front 4 Replacing rear panel Insert the...
Page 18: ...nection socket Detector 1 NOTICE Recommendation In AnCat systems with 2 detectors anions to Detector 1 cations to Detector 2 2 Connecting the detector outlet capillary Guide the detector outlet capil...
Page 19: ...C page 12 With instruments with sequential suppression to the MCS see Con necting the detector inlet capillary to the MCS page 12 NOTICE In order to prevent unnecessary peak widening after separation...
Page 20: ...2744 070 short pressure screws 5 4 1 2 3 4 Figure 5 Connector detector MSM 1 Detector inlet capillary 2 MSM MSM HC outlet capillary labeled with out 3 Coupling 6 2744 040 4 Pressure screws short 6 27...
Page 21: ...spect that moisture has gotten inside the instrument Only personnel who have been issued Metrohm qualifications may perform service and repair work on electrical and electronic parts Connecting the po...
Page 22: ...Detector 3 Start up The IC Conductivity Detector is put into operation together with the IC instrument in which it has been inserted to Additional information can be found in the Start up chapter in t...
Page 23: ...ust be set to 5 MPa in MagIC Net If the conductivity detector is blocked you should first of all check whether the blockage results from the capillary ends being pressed together too firmly In this ca...
Page 24: ...ressure in the system markedly increases Conductivity detector blocked see chapter 4 page 15 Shorten capillary ends by several mm Rinse detector against the normal flow direction Conductivity detec to...
Page 25: ...ft 0 2 nS cm per hour Measuring rate 10 measurements per second for optimum results without filtering Resolution 0 0047 nS cm Baseline Noise 0 2 nS cm typical for sequential suppression Conductivity d...
Page 26: ...igh density 6 3 Energy supply Power supply via detector cable 5 V 6 5 V 42 V 6 4 Reference conditions Ambient tempera ture 25 C 3 C Instrument status 40 min in operation 6 5 Dimensions Width 108 mm He...
Page 27: ...m com into your Internet browser 2 Enter the article number e g 2 850 9010 into the search field The search result is displayed 3 Click on the product Detailed information regarding the product is sho...
Page 28: ...acing 7 Detector cable Connecting 10 Detector inlet capillary 2 Connecting 11 Detector outlet capillary 3 Connecting 10 Dimensions Technical specifications 18 E Electrostatic charge 5 Energy supply Te...