3 Installation
850 Professional IC–Anion – MSM-HC – MCS – LP Gradient
■■■■■■■■
39
Table 1
Smallest possible eluent percentage depending on installed
pump head and set flow
Flow
Standard pump head
Macro pump head
1 mL/min
10 %
2.5 %
2 mL/min
20 %
5 %
3 mL/min
not recommended
7.5 %
Depending on the application, an optimal assignment of each eluent to a
valve is documented in the Application Notes. To obtain a reproducible
gradient profile, this eluent-valve assignment must remain unchanged.
Each eluent is degassed by means of its own eluent degasser
1
2
3
4
6
5
7
8
9
Figure 18
Low pressure gradient
1
Mixing valve
2
Eluent degasser
for Eluent B
3
Eluent degasser
for Eluent C
4
Connection tubing 6.1834.100
Connects eluent degasser
and mixing
valve C. Pre-installed.
5
Eluent inlet C
6
Connection tubing 6.1834.100
and the
mixing valve input B. Pre-installed.
7
Eluent inlet B
8
Eluent inlet A
9
Eluent outlet