Chapter II
Benchtop multimeter
26
5.5.
Resistance / Continuity
Ranges
Accuracy
Measurement
current
Protection*
Resolution
500
Ω/€
0.1% R + 5 D
1 mA
600 V
RMS
10 m
Ω
5
k
Ω
0.1% R + 3 D
100 µA
600 V
RMS
100 m
Ω
50 k
Ω
0.1% R + 3 D
10 µA
600 V
RMS
1
Ω
500 k
Ω
0.1% R + 3 D
1 µA
600 V
RMS
10
Ω
5 M
Ω∗∗
0.3% R + 3 D
100 nA
600 V
RMS
100
Ω
50 M
Ω∗∗
1% R + 3 D
10 nA
600 V
RMS
1 k
Ω
* Overload protection can be reset automatically.
**It is highly recommended to use very short and shielded test leads for measurements in
this range (>1M
Ω).
When performing resistance measurements, connect the COM terminal to the earth.
Number of points:
50 000 (or 5 000 - refer to § 3.3.)
Range selection:
Automatic or manual
(fixed in continuity mode)
Maximum open circuit voltage:
7 V
Detection threshold in continuity mode:
10
Ω
to 20
Ω
Response time in continuity mode:
1 ms
5.6. Capacitance
"
Note
Discharge all capacitors before taking measurements.
Ranges
Accuracy
Measurement
current
Max measurement
time
Protection*
Resolution
50 nF**
1.5% R + 2 D
100 nA
0.5 s
600 V
RMS
10 pF
500 nF
1% R + 2 D
1 µA
0.5 s
600 V
RMS
100 pF
5 µF
1% R + 2 D
10 µA
0.5 s
600 V
RMS
1 nF
50 µF
1% R + 2 D
100 µA
0.5 s
600 V
RMS
10 nF
500 µF
1% R + 2 D
1 mA
1.5 s
600 V
RMS
100 nF
5 000µF
2% R + 2 D
1 mA
3 s/mF
600 V
RMS
1 µF
50 mF
2% R + 2 D
1 mA
3 s/mF
600 V
RMS
10 µF
* Overload protection can be reset automatically.
**It is highly recommended to use very short and shielded test leads for measurements in
this range.
Number of points:
5 000
Range selection:
Automatic or manual
Maximum open circuit voltage:
7 V
5.7.
Diode threshold voltage measurement
Measurable voltages:
0 to 2 V
Measurement current:
1 mA ± 20%
Resolution:
1 mV
Protection:
600 V
RMS
, can be reset automatically