Chapter II
5000-count portable digital multimeter, with current input
24
5.2.3. DC current
Selector switch
position
Ranges
Accuracy
Max voltage
drop
Protection
Fuses
(∗)
Resolution
500 mA
DC
500 mA
DC
0.3 %R + 2 D
< 800 mV
600 V
RMS
F1 + F2
100 µA
DC
10 A
DC
(∗∗)
10 A
DC
1 %R + 5 D
< 700 mV
600 V
RMS
F2
10 mA
DC
(∗)
Refer to fuse specifications section 3.5.3.
(∗∗)
Acceptance 20 A overload during 30 s. max, observing a 5 min break at least
between two measurements
Number of points :
5 000
«
OL
» on the display (range 500 mA
DC
only) and intermittent signal sounds in case of
range overpassing.
5.2.4. AC currents (AC+DC)
Selector switch
position
Ranges
Accuracy
40 Hz to 1 kHz
Max voltage
drop
Protection
Fuses
(∗)
Resolution
500 mA
AC+DC
500 mA
AC+DC
1.5 %R + 2 D
< 800 mV
600 V
RMS
F1 + F2 100 µA
AC+DC
10 A
AC+DC
(∗∗)
10 A
AC+DC
2.5 %R + 5 D
< 700 mV
600 V
RMS
F2
10 mA
AC+DC
(∗)
Refer to fuse specifications section 3.5.3.
(
∗∗
)
Acceptance 20 A overload during 30 s. max, observing a 5 min break at least between
two measurements
Specifications applicable from 5 % to 100 % of the range for a sinusoidal signal in band
40 Hz to 1 kHz.
Number of points :
5 000
Additional error according to crest factor :
1 % for a crest factor of 1.5 to 2
4 % for a crest factor of 2 to 3
«
OL
» on the display (range 500 mA
AC+DC
only) and intermittent signal sounds in case
of range overpassing.
5.2.5. Resistance / Continuity test
Selector switch
position
Ranges
Accuracy
Measurement
current
Protection
(∗)
Resolution
500
Ω
0.3 %R + 3 D
1 mA
0.1
Ω
500
Ω
1 mA
0.1
Ω
5 k
Ω
0.3 %R + 3 D
100 µA
1
Ω
Ω
50 k
Ω
10 µA
600 V
RMS
10
Ω
500 k
Ω
1 µA
100
Ω
5 M
Ω
0.5 %R + 3 D
100 nA
1 k
Ω
50 M
Ω
1 %R + 5 D
50 nA
10 k
Ω
(∗)
Overload protection can be reset automatically.
Number of points :
5 000
Range selection :
automatic or manual (fixed in continuity
mode)
Maximum open circuit voltage :
4 V
Detection threshold in continuity mode :
10
Ω
to 20
Ω
Response time in continuity mode :
< 1 ms
For measurements in 5 M
Ω
and 50 M
Ω
ranges, the use of shielded connection is highly
recommended.