IX 3131B
II - 39
Capacitance measurement
100 Hz
120 Hz
1 kHz
10 kHz
Range
Parallel Series Parallel Series Parallel Series Parallel Series
10 mF
√
√
√
√
1000
µ
F/1 mF
√
√
√
√
√
√
200/50
µ
F
√
√
√
√
√
√
√
√
20
µ
F
√
√
√
√
√
√
√
√
2000 nF
√
√
√
√
√
√
√
√
200 nF
√
√
√
√
√
√
√
√
20 nF
√
√
√
√
√
√
√
√
2000 pF
√
√
√
√
200 pF
√
√
Inductance measurement
100 Hz
120 Hz
1 kHz
10 kHz
Range
Series Parallel Series Parallel Series Parallel Series Parallel
1000 H
√
√
√
√
200/100 H
√
√
√
√
√
√
20 H
√
√
√
√
√
√
2000/1000 mH
√
√
√
√
√
√
√
◊
200 mH
√
√
√
√
√
√
√
√
20 mH
√
◊
√
◊
√
√
√
√
2000
µ
H
√
√
√
√
3.4.11. Calibration
Calibration is available in all ranges. Simply press and hold « CAL » key for one
second to enter the calibration mode and calibration prompts will be displayed.
•
Follow the prompts for open connection (
) or short connection (
) the
« CAL » button.
•
After calibration is completed, the meter will be restored to normal display and
ready for normal usage.
The calibration will save the internal parameter as well as external connector
residues for further measuring. It is highly recommended to calibrate extremely high
or low ranges for L, C and R before making precision measurements. Calibration
prompts will be displayed automatically every time those ranges are manually or
functionally selected, (e. g. REL, TOL, MIN / MAX / AVG etc.). Simply follow the
open connector (
) or short connector (
) instruction and then press the
« CAL » button. You may skip the calibration by pressing the « D / Q /
θ
» button.
Note
Changing measurement frequencies is handled the same as
selecting a different hardware range, and so automatic
calibration prompts will be displayed in the recommended
ranges.
Summary of Contents for IX 3131B
Page 24: ...IV 24 IX 3131B IX 3131B BENCH TYPE L C R METER Operating manual Chapter II ...
Page 47: ...IV 47 IX 3131B IX 3131B RLC MESSBRÜCKE Bedienungsanleitung Kapitel III ...
Page 71: ...IV 71 IX 3131B IX 3131B PONTE RLC Manuale di Funzionamento Capitolo IV ...
Page 94: ...IV 94 IX 3131B IX 3131B PUENTE RLC Manual de funcionamiento Capítulo V ...