![Metrix DOX 2 Series User Manual Download Page 26](http://html1.mh-extra.com/html/metrix/dox-2-series/dox-2-series_user-manual_1779215026.webp)
24
Kit Demo for METRIX Oscilloscopes
Demo:
with:
DOX2025
DOX2040
DOX2100
DOX2xxx
Test Signal
no. 12: Recorder-5 signals
Nature
Tracking of 5 slow signals having varied shapes and characteristics
Specs
Duration of each signal
1s, amplitude 1.5V < Vpp < 3.5V
Oscilloscope Settings
Duration-2s scale - 40µs-MAIN = 500mV/div., DC coupling
Trigger
None at first, then threshold(s) on MAIN, level according to signal
Modes
"Source/Level" triggering,
Objectives
Elementary presentation of the "Scan" mode for the < 50ms/div time base
Attention: the Scan and LongMem modes, along with
the Scan and Delayed modes, are not compatible
Attention: in Scan mode, for the "Level" to be active, the triggering must be in
Normal mode when the Type of Trigger is on "Front"
Observe in Scan mode the signals delivered by the HX0074 demonstrator:
Using triggering on pulse width and acting on the position of the level, it will be possible to trigger on each of the
5 slow signals.
By placing the triggering level close to zero and programming the pulse width, it is possible to synchronize to
the damped sine wave, the low pulse, and the low ramp.
Summary of Contents for DOX 2 Series
Page 1: ...X04128A02 Ed 1 09 15 Demonstration SET for METRIX DOX 2xxx Oscilloscopes User s guide...
Page 9: ...Kit Demo for METRIX Oscilloscopes 7 FFT of the signal on channel CH2...
Page 16: ...14 Kit Demo for METRIX Oscilloscopes Zoom by 1000 in LongMem mode...
Page 27: ...Kit Demo for METRIX Oscilloscopes 25...