Technical characteristics of the MTX 3290
6,000- and 60,000-count digital multimeters
41
Technical characteristics of the MTX 3290 (continued)
V
AC+DC TRMS
Protection: 850Vpk
Range
Operating
range
Specified
measure.
range
3)
Resolution
Uncertainty
DC (
±
)
Uncertainty
AC (
±
)
Additional
uncertainty
F (Hz)
1)
Pass
band
Input
impedance
// <50 pF
Peak
factor
600mV
0 to
600.0mV
60.0 to
600.0mV
0.1mV
0.8% L
±10D
2% L
+ 0.18%x
[F(kHz)-1]L
±5D
45<F<65Hz
0.3% L
typ.
at 100Hz
0 .7% L
typ.
at 150Hz
1.8% L
typ.
at 300Hz
30% L
typ.
10Hz
to
20kHz
10.9M
Ω
3 to
500mV
6V
0 to
6.000V
0.600 to
6.000V
0.001V
2% L
+ 0.18%x
[F(kHz)-1]L
±3D
10Hz
to
20kHz
10.9M
Ω
3 to
5V
60V
0 to
60.00V
6.00 to
60.00V
0.01V 10.082M
Ω
3 to
50V
600V
2)
0 to
600.0V
60.0 to
600.0V
0.1V 10.008M
Ω
3 to
500V
1) See the typical curve of the 300Hz filter.
2) The LCD indicates “+OL” above +620V, “-OL” below -620V or above 620VRMS.
3) From 1kHz, the measurement must exceed 15% of the range.
Secondary measurements and displays: FREQ (AC coupling),MAX,MIN,AVG,PEAK
V
LowZ AC
Protection: 850Vpk
The pass band is reduced to 300Hz if the filter is activated. The frequency
measurement is made like the measurement in a 300Hz pass band.
Range
Operating
range
Specified
measurement
range
3)
Resolution Uncertainty (
±
)
Additional
uncertainty F
(Hz)
1)
Input
impedance
//<50 pF
Peak
factor
600mV
0 to
600.0mV
60.0 to
600.0mV
0.1mV
2 .2% L+
0.25%x
[F(kHz)-1]L
±5D
45<F<65Hz
0.3% L typ.
at 100Hz
0.7% L typ.
at 150Hz
1.8% L typ.
at 300Hz
30% L typ.
≅
300k
Ω
3 to
500mV
6V
0 to 6.000V
0.600 to
6.000V
0.001V
2 .2% L+
0.18%x
[F(kHz)-1]L
±3D
3 to 5V
60V
0 to 60.00V
6.00 to
60.00V
0.01V
3 to 50V
600V
2)
0 to 600.0V
60.0 to
600.0V
0.1V
3 to 500V
1) See the typical curve of the 300Hz filter.
2) The LCD indicates “+OL” above +620V, “-OL” below -620V or above 620VRMS.
3) From 1kHz, the measurement must exceed 15% of the range.
Secondary measurements and displays: FREQ (AC coupling), MAX, MIN, AVG,PEAK