To ensure the accuracy of the instrument and prevent it from damage, please pay attention to the maintenance. 
Some points for attention are hereby given: 

6.1

 The instrument should be placed in a dry and well-ventilated room, so as to prevent the optical components 

from becoming damp and going moldy. 

6.2

 After measuring corrosive liquid, it it necessary to do cleaning without delay (including optical components, 

metal parts and painting surface), so as to prevent the instrument from corrosive damage. 

6.3

 No solid impurityis permitted to exist in the sample to be measured. When measuring the solid sample, 

prevent the working surface of the refracting prisms from being roughed and scratched. 

6.4

 Regularly keep the instrument clean. Strictly prohibit hands wet with oil or perspiration from touching the 

optical components. If the surface of optical components is contaminated by dirt, use high-grade chamois or 
long-fiber absorbent cotton to gently wipe away and blow out with leather blower. If the surface of optical 
components is contaminated by greasy dirt, clean out with the mixed solution of alcohol and diethy1 ether 
without delay. 

6.5

 The instrument should avoid violent vibrations or shock, to prevent the optical components from being 

damaged, and keep the accuracy of the instrument as well. 

Summary of Contents for RST111

Page 1: ...ABBE Analogic refractometer RST111 Please read the User Manual carefully before use and follow all operating and safety instructions user manual english...

Page 2: ...are representative with the following information Serial number Description of problem Your contact information Warranty This instrument is warranted to be free from defects in materials and workmansh...

Page 3: ...perature between 0 and 70 Both the refractive index and the average dispersion are the important optical constants through which we can understand the optical performance purity concentration as well...

Page 4: ...0 Hand wheel 16 Adjusting cavity 5 Light entering prism holder 11 Refraction prism holder 17 Casing 6 Dispersion adjusting wheel 12 Condensing lens 18 Adapter connector for 4 thermostats 1 Light enter...

Page 5: ...and 11 the refracting prism holder linked by the rotation shaft 2 the light entering prism can be opened or closed When the two prism holders are tightly closed and locked by the hand wheel 10 a unif...

Page 6: ...x in the field of view of the eyepiece the operating methods of sighting and reading are the same as above 5 2 3 Measurement of translucent solid Stick the polished surface of the translucent solid to...

Page 7: ...in the sample to be measured When measuring the solid sample prevent the working surface of the refracting prisms from being roughed and scratched 6 4 Regularly keep the instrument clean Strictly pro...

Page 8: ...8...

Page 9: ...ake measuring the average dispersion of distilled water as an example When temperature is 20 nD 1 3330 The reading of dispersion value indicated on the scale ring will be Turn in a certain direction T...

Page 10: ...tomatically Turn off and put the key for four seconds turn on put the key back on When the temperature is higher than 70 C display will show H when is lower than 50 C display will show L C When the ba...

Page 11: ...desecharse en vertederos De conformidad con la Directiva 2002 96 CE los usuarios de la Uni n Europea de aparatos el ctricos y electr nicos tienen la oportunidad de retornar el instrumento para su eli...

Page 12: ...www labbox com...

Reviews: