MI 2892 Power Master
Operating the instrument
83
dMax
∆Umax - maximum absolute difference between any of the U
Rms(1/2)
values during the RVC event and the final arithmetic mean 100/120 U
Rms(1/2)
value just
prior to the RVC event.
dUss
∆Uss - is the absolute difference between the final arithmetic mean
100/120 U
Rms(1/2)
value just prior to the RVC event and the first arithmetic mean
100/120 U
Rms(1/2)
value after the RVC event.
Table 3.75: Keys in RVC Events table group view screens
STAT
Shows event statistics (phase by phase).
RVC
Returns to RVC Events table group view screen.
Selects RVC Event.
Returns to RVC Events table group view screen.
Returns to “RECORDERS” submenu.
3.20
Memory List
Using this menu user can view and browse saved records. By entering this menu, information about
records is shown.
Figure 3.59: Memory list screen
F4