MI 3290 Earth Analyser
Appendix F - The balancing procedure
156
Figure E.7: Example of Search result within Auto Sequence
® group
V.
Elements of an Auto Sequence
®
Auto Sequence
® steps
There are three kinds of Auto Sequence
® steps.
Header
The Header step is empty by default.
Flow commands can be added to the Header step.
Measurement step
The Measurement step contains a Single test and the Operation after end of test flow command
by default. Other Flow commands can also be added to the Measurement step.
Result
The Result step contains the Result screen flow command by default. Other Flow commands
can also be added to the Result step.
Single tests
Single tests are the same as in Metrel ES Manager Measurement menu.
Limits and parameters of the measurements can be set. Results and sub-
results can’t be set.
Flow commands
Flow commands are used to control the flow of measurements. Refer to chapter
Number of measurement steps
Often the same measurement step has to be performed on multiple points on the device under
test. It is possible to set how many times a Measurement step will be repeated. All carried out
individual Single test results are
stored in the Auto Sequence® result as if they were
programmed as independent measuring steps.
VI.
Creating / modifying an Auto Sequence
®
If creating a new Auto Sequence
® from scratch, the first step (Header) and the last step (Result)
are offered by default. Measurement steps are inserted by the user.