MI 3103 GigaOhm Specifications
18
Rcon
Measuring range R
(
)
Resolution
(
)
Accuracy
0.0
199.9
0.1
(5 % of r 3 digits)
200
999
1
1000
1999
1
10 % of reading
Open-terminal test voltage: 4 V - 7 V DC
Compensation of test leads (up to 5
): yes
Short-circuit test current: up to 7 mA
Measurement mode:
continuous measurement
One set of alkaline cells can supply the instrument for more than 90 hours.
Voltage U
Measuring range U
(V)
Resolution
(V)
Accuracy
0
600
1
(3 % of r 3 digits)
Nominal frequency range: DC, 16 Hz – 450 Hz
Measurement mode:
continuous measurement, TRMS
Measuring range in operating conditions.