Multinorm/FonS/Poly
18
3.1.5 Displaying single parameter
First select the desired parameter with the
and keys. Then press the F3
(
Single
) key to enable this function which allows you to focus on only one measurement
parameter. All values of this single parameter are displayed (
Min
,
Avg
,
Max
,
Cur
,
Start
Time
,
Stop Time
). Press the F3 (
All
) key to return to display of all measurement
parameters. Picture 3.7 shows single parameter displays for air temperature and
illumination parameters.
Picture 3.7: Single measurements
3.1.6 Save function
Press the F4 (
Save
) key to enter
Save menu
(described in chapter 4.2.1) which
allows you to store measurements. If measurement is not already stopped, pressing the
F4 (
Save
) will automatically stop the measurement.
3.2 Logger mode (Measurements)
3.2.1 Overview
This measuring mode is intented to collect, handle and periodicly save
measuring data in non volatile memory from all probes that are connected to different
instrument measuring ports. Time period between two memorized measurements can
be set up between 2 s and 60 min. This data can be downloaded to PC by A1166
LabLink PRO/ A1134 SensorLink PRO (see corresponding user manual). For proper
probe connection see the user manual for desired probe.
3.2.2 Description
3.2.2.1 Entering logger mode
In the normal measuring mode change the function keys to second set by
pressing the
key. Then press the F1 (
Logger
) key. Instrument is now in logger mode,
caption
LOG
appears in the status window.
Logging procedure saves measurement records into non volatile memory. To
define (change) location levels of the logging measurement records press the F4 (
Loc
)
key.