MI 3102(H) BT
Eurotest
XE
Measurements
64
5.11
Harmonics
Harmonics are components of the voltage and currents signal with an integer multiple of the
fundamental frequency. The harmonic values are an important parameter of power quality.
selection for instructions
on key functionality.
Figure 5.48: Harmonics menu
Settings and parameters in Harmonics function
Input
Displayed parameters [ voltage U or current I]
h:0.-.h:11
Selected harmonic
Connection for the Harmonics measurement
See
Harmonics measurement procedure
Select
HARMONICS
function from the
OTHERS menu
.
Select test
parameters
(optional).
Connect
voltage test leads and current clamp to the instrument.
Connect
the voltage test leads and current clamp to the item to be tested (see
Press the
TEST
key to start the continuous measurement.
Press the
TEST
key again to stop the measurement.
Store
the result by pressing the
MEM
key (optional)
Figure 5.49: Examples of Harmonics measurement results
Displayed results for the Harmonics measurement:
Uh
............ TRMS voltage of selected harmonic
Ih
.............. TRMS current of selected harmonic
THDU
....... voltage total harmonic distortion
THDI
......... current total harmonic distortion
Notes:
For setup of current clamps (see chapter
Parameters (input and number of harmonic) can be changed and results can also be
stored while the measurement is running.
Displayed graph is auto-ranged.