MI 3152(H)
Eurotest
XC (2,5 kV)
Tests and measurements
99
7.5
R low – Resistance of earth connection and
equipotential bonding
Figure 7.18: R low measurement menu
Measurement parameters / limits
Output
1)
[LN, LPE]
Bonding
[Rpe, Local]
Limit(R)
Max. resistance
[Off, 0.05
Ω ... 20.0 Ω]
1)
R low measurement depends on Output parameter setting, see table below.
Output
Test terminals
LN
L and N
LPE
L and PE
Connection diagram
Figure 7.19: Connection of 3-wire test lead plus optional Extension lead