A 1632
eMobility Analyser
General description
5
1.1.1 Markings on the instrument:
Read the Instruction manual with special care to safety operation«. The
symbol requires an action!
Mark on your equipment certifies that it meets requirements of all subjected EU
regulations
.
This equipment should be recycled as electronic waste.
This equipment is protected by reinforced insulation.
1.1.2 Notes related to measurement functions
R iso
The resistances in OUTPUT area, between terminals L/L1-PE, L2-PE, L3-PE are ca 100
MΩ and between terminals L/L1-N, L2-N, L3-N, Lx-Ly are ca 200 MΩ. This should be
considered if Riso is measured.
The OUTPUT voltage LED indicators can light on during the insulation test. This has no
particular meaning and has no influence on the measurement.
Diagnostic test (EVSE)
Because of capacitive crosstalk between conductors in the three phase Output adapters
a certain voltage U
LxN
will be displayed on not connected phases.
Do not use A 1631 Monitor adapter cable for charge currents higher than 32 A.