MI 2087 AL2 EasiTEST
Other operations
33
4 Other operations
4.1 Reset of the instrument
If any malfunction is noticed when dealing with the EasiTest, it is advisable to
RESET
the instrument. In that case all settable parameters will be set to their initial values;
see the table below. Stored results will not be erased.
How to reset the instrument
q
Switch OFF the instrument.
q
Press the
CAL
key and keep it pressed, while switching ON the instrument.
“
rES
” message will be displayed for a while meaning RESET function has been
completed.
Parameter
Function
Initial value
Compensation of test leads
Continuity
Annulled
Nominal differential current
All RCD functions
30 mA
Type of RCD
All RCD functions
G (general)
Limit contact voltage
All RCD functions
50 V
Multiplier of nominal
differential current
t
∆
N
×1/2
Test voltage
Insulation
250 V
Table 2:
Initial values of settable parameters