background image

A 1740 Calibration Box 5kV 

A 1740 Description / A 1740 Application

 

 

2.3  Standards applied 

 
A 1740 Calibration Box 5kV is manufactured and tested according to the following 
regulations, listed below. 
 

Safety 

 

EN 61010 - 1 

Safety requirements for electrical equipment for measurement, 
control, and laboratory use 

– Part 1: General requirements 

 

A 1740 Description 

 

Outlook of calibration unit 

 

   

          

Figure 3.1: 

Calibration Box 5kV (A 1740)

 

Legend: 

1...................................... 10 M

 connection stub, (intended for connection of +Rx) 

2...................................... 100 M

 connection stub, (intended for connection of +Rx) 

3...................................... 1 G

 connection stub, (intended for connection of +Rx) 

4...................................... 10 G

 connection stub, (intended for connection of +Rx) 

5...................................... COMMON connection stub, (intended for connection of 

–Rx) 

 

A 1740 Application 

 

Procedure:

 

 

Connect 

–RX to COMMON connection stub and +Rx to 10 M

 connection 

stub.

 

 

Set test voltage, press 

START / STOP

 key on the instrument and wait until 

result becomes stable, then verify indication.

 

 

Press 

START / STOP

 key again to stop the measurement.

 

 

Repeat the same procedure with the rest test values.

 

 
The instrument does not require any adjustment if indications are inside the stated 
limits. 

 