12
Power Supply: 3 V standard button battery x 1 (IEC-CR2032; ANSI-NEDA-
5004LC)
Power Consumption (typical): 2 mA
APO Consumption (typical): 2.2 µA
APO Timing: Idle for 3 minutes
Dimension / Weight
L 113 mm x W 53 mm x H 10.2 mm / Approx. 78 gm
Special Features
AutoTect
(Automatic V and
Ω
selection) and VolTect
(Electric Field
Detection)
Agency Approvals
P >
Safety: Meets IEC61010-1, UL61010B-1, CAN/CSA-C22.2 No. 1010.1-92,
CAT II 450 V and CAT III 300V, Pollution Degree 2, Class 2
E.M.C. Meets EN61326 (1997, 1998/A1), EN61000- 4-2 (1995), and
EN61000-4-3 (1996). This product complies with requirements of the
following European Community Directives: 89/ 336/ EEC (Electromagnetic
Compatibility) and 73/ 23/ EEC (Low Voltage) as amended by 93/ 68/ EEC
(CE Marking). However, electrical noise or intense electromagnetic fields in
the vicinity of the equipment may disturb the measurement circuit.
Measuring instruments will also respond to unwanted signals that may be
present within the measurement circuit. Users should exercise care and
take appropriate precautions to avoid misleading results when making
measurements in the presence of electronic interference.
Accessories
Silicon holster, Battery installed, and User’s manual
Optional Accessories
VC3 soft carrying case with zipper
Electrical Specifications
(Accuracy @ 23
°
C +/- 5
°
C and < 75% R.H.)
RF Field @ 3 V/m: Specified ac 45 d (Capacitance not specified)