15
AC Voltage
Range Accuracy
50 Hz – 60 Hz
6.000 V, 60.00 V, 450.0 V, 600 V
±
(1.5%+5 dgt)
CMRR: > 60 dB @ dc to 60 Hz, Rs=1 k
Ω
Input Impedance: AutoTect
Lo-Z V ac: 160 k
Ω
, 160 pF nominal
Hi-Z ACV: 5M
Ω
, 90pF nominal
ACV AutoTect
Threshold: > 1.5 V ac (50 Hz/60 Hz) nominal
Capacitance
Range
1
Accuracy
2
100.0nF, 1000nF, 10.00
µ
F,
100.0
µ
F
3)
±
(3.5%+6 dgt)
4
1)
Accuracy below 50nF is not specified
2)
Accuracies with film capacitor or better
3)
Updates > 1 minute on large values
4)
Specified with battery voltage above 2.8 V (half full battery). Accuracy
decreases gradually to 12% at low battery warning voltage of approx.
2.4 V.