Maintenance and Troubleshooting
|
5-25
When is an Auto-Setup Required / Not Required?
The Auto-Setup optimizes PMT (photomultiplier tube) gains and the
source lamp currents. After the Auto-Setup is complete, the Automatic
Lamp Control (
Alc On
) is automatically turned on. View the current sta-
tus of this function on the
Alc
screen (‘
0
’ = Alc Off; ‘
1
’ = Alc On).
Auto-Setup is required when:
• The “
Warning PMT Signal
” alarm is displayed under
Error Condition
on the
Status
tab.
• One or both source lamps are replaced or adjusted.
• The Measuring Cell Windows and/or optics are cleaned.
• Any optical filter is replaced.
• One or both PMTs are replaced.
• The
Optical Bench
board is replaced.
Auto-Setup Completion Number
At the end of every Auto-Setup the analyzer displays a completion num-
ber that is the product of the lowest and highest transmittance. A normal
completion number is 1.00, ±0.25.
IMPORTANT
Although a good completion number is indicative of a successful
Auto-Setup, do not rely on it solely.
Always check the PMT Level and PMT Balance after every Auto-
Setup. See “PMT Level and PMT Balance,” following this section, for
details on acceptable levels of these two parameters.
If the completion number is outside of this range it is an indication that
the
Measure
and
Reference
signals are not at optimum ratios to each other.
The most common cause of an abnormal completion number is initiating
Auto-Setup when the Measuring Cell has not been properly purged. This
results in completion numbers
lower
than normal. The most likely cause
of a
higher
than normal completion number is incorrect placement of the
PMTs in their respective sockets. Although the sensitivity of each PMT is
similar, the
Measure
path PMT typically has a slightly higher gain than the
Reference
path PMT. Swapping the PMTs from the factory-set positions can
produce higher than normal completion numbers.
NOTE
(Alc) CFG
F2 7
Del 0 Ent
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...