Maintenance and Troubleshooting
|
5-3
If the analyzer sets alarms that indicate a faulty component requires
replacement (see “Troubleshooting and Diagnostics” in this chapter),
follow the corresponding replacement procedure in this chapter.
For complex maintenance procedures not discussed in this manual,
such as replacing heaters, RTDs, or electronic boards, contact
AMETEK. Or, review Chapter 6 and then contact AMETEK for as-
sistance with returning the assembly/analyzer to the factory for repair.
Analyzer Preventive Maintenance Schedule
To reduce the occurrence of problems with the analyzer follow this sched-
ule, which lists the frequency of maintenance required when caring for
the analyzer. Since most analyzer problems originate within the sample
system, the primary objective of this schedule is proper care of the sample
system.
Frequency Task
Daily
Check for Warning or Fault Alarms
Check the top-right corner of the User Interface for the
S
character. If
S
appears, view the
HStatus
and
MStatus
screens for current alarms
to help isolate the problem. See “Troubleshooting and Diagnostics” in
this chapter for alarm messages and corrective action.
Check the history buffers (
HCHist
and
MCHist
) for recurring alarms.
The history buffers contain alarms that have been reset.
Monthly
Check the PMT Signals
Manually Zero the analyzer. (See “Manually Zeroing the Analyzer” in
Chapter 3.)
View the Show Signals (
SIG
) screen and record the analyzer PMT
signals. All signals should return to within 5 % of the values recorded
the previous month. If the
Measure
signals show significant loss, the
Measuring Cell may be contaminated. Cleaning the Measuring Cell
window can be delayed until the signals are less than 50 % of their
“clean” measured value. Refer to your log book to examine monthly
recordings.
Immediately following this monthly maintenance, continue with the
monthly task “Check Analyzer Response Time” below.
!
WARNING
(HStatus) RUN
F5 4
1..7
(MStatus) RUN
F5 5
1..7
(HCHist) RUN
F5 •
1..9
(MCHist) RUN
F5 –
1..9
(SIG) RUN
F6 1
1..6
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...