Controller / User Interface | 4-41
Analyzer Control Mode (Automatic Control by Analyzer)
When the analyzer’s Flow Control mode is set to
Analyzer Control
, the
analyzer automatically determines and sets the mode of operation by
monitoring the state of the Fault alarm relay contacts.
If there are no active Fault alarms, the analyzer will operate in
Analyzer
Control
mode (
automatic
Sample
). When in
automatic
Sample
mode, there
is no indicating character on the User Interface.
If the analyzer detects a Fault alarm while it is operating in Automatic
Sample mode,
S
is displayed on the top-right line of the User Interface.
When this occurs, the analyzer will switch to
automatic
Backpurge
, and
“
b
” is displayed on the top-left line (as opposed to a “
B
”, which indicates
manual
Continuous Backpurge
). The analyzer will not switch back to
automatic
Sample
mode until the cause of the alarm has been corrected
(cleared).
If a fault alarm is triggered, the analyzer will switch from
automatic
Sample
mode (sample gas flowing) to
automatic
Backpurge
mode
(zero gas flowing). If the alarm automatically clears (resets) itself, the
analyzer will switch back to
automatic
Sample
mode. If the analyzer
is in
automatic
Backpurge
mode, sample gas stops flowing and the
sample system is flushed with Zero gas until the alarm has been man-
ually corrected, at which time the analyzer switches back to
automat-
ic
Sample
mode. If this condition recurs, refer to
Troubleshooting
and Diagnostics
in Chapter 5 and review the corrective action for
the corresponding alarm, for information on how to troubleshoot the
condition.
Because a low temperature is defined as a fault, the analyzer will switch to
automatic
Backpurge
mode for the length of time required for all tempera-
ture zones to stabilize at their Set Points.
To set the analyzer’s Flow Control mode to
Analyzer Control
mode and
return to CAL mode normal display, press:
CAL>F5 2 Del 0 Ent Ent
NOTE
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...