Controller / User Interface | 4-33
CAL Mode – F1 Commands
Keystroke Term
Definition
F1 •
—
—
F1 –
—
—
F1 0
Auto/Zero?
Automatic adjustment of the analyzer Zero. The Zero gas (air) solenoid is turned on and off
automatically during the Zero. The Zero Offset is adjusted based on the average reading
during the last 25 % of the
Timer0
countdown only if
AdjDisable
is set to ‘
0
’.
F1 y
Auto/Span
y
? Allows you to initialize the analyzer to automatically introduce Calibration gas into the
sample system, where ‘
y
’ =
F1 1..F1 2
.
F1 1
= Auto/Span1?
F1 2
= Auto/Span2? –
Used only with O
2
Sensor (General Purpose applications)
F1 2
= Auto/Span2? –
System Isolation Blow-Back Solenoid (Optional)
Auto/Span
Only the Span function for ‘
y
’ is performed. The Calibration gas solenoid valve connected to
SSR output ‘
y
’ is turned on and off automatically during the sequence.
The analyzer calibration is automatically adjusted based on the average reading during the
last 25 percent of the countdown after a
Yes
response only if
AdjDisable
is set to ‘
0
’.
If a multi-component calibration gas mixture is used, the spans for all components in the
mixture are done simultaneously.
Isolation Blow-Back Solenoid (Optional)
This system is used in place of the “AutoSpan2?” function on systems requiring Sample
Line Isolation and Blow-Back Solenoid (custom configurations).
F1 3..
—
—
F1 4
—
—
F1 5
—
—
F1 6
—
—
F1 7
PrbBlowBk? Used on systems that use a sample probe with a Blow-Back function. When you press
F1 7
from this menu, “
PrbBlowBk?
” is displayed. Press
Ent
to initiate a manual probe
Blow-Back.
F1 8
—
—
F1 9
Auto/Cal?
Allows you to initialize the analyzer to automatically cycle sequentially through the Auto-
Zero and Auto-Span functions. Only the functions with timers set to non-zero durations are
included.
The Zero Offset and calibration will be automatically adjusted only if
AdjDisable
is set to ‘
0
’.
Summary of Contents for 920
Page 39: ...Specifications 2 11 ...
Page 40: ...2 12 Model 920 Multi Gas Analyzer ...
Page 41: ...Specifications 2 13 2019 02 05 ...
Page 42: ...2 14 Model 920 Multi Gas Analyzer ...
Page 43: ...Specifications 2 15 ...
Page 44: ...2 16 Model 920 Multi Gas Analyzer ...
Page 45: ...Specifications 2 17 ...
Page 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Page 48: ...2 20 Model 920 Multi Gas Analyzer ...
Page 49: ...Specifications 2 21 ...
Page 50: ...2 22 Model 920 Multi Gas Analyzer ...
Page 51: ...Specifications 2 23 ...
Page 52: ...2 24 Model 920 Multi Gas Analyzer ...
Page 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Page 55: ...Specifications 2 27 ...
Page 56: ...2 28 Model 920 Multi Gas Analyzer ...
Page 57: ...Specifications 2 29 ...
Page 58: ...2 30 Model 920 Multi Gas Analyzer ...
Page 59: ...Specifications 2 31 ...
Page 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Page 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Page 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Page 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Page 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Page 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Page 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Page 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Page 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Page 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Page 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Page 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Page 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Page 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Page 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Page 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Page 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...