Operation
MicroVol-1100 User manual 1.6
Up to six on and off periods can be programmed into the timer. If the current
time is within any of the six programmed periods, then the Time condition
will be satisfied. Note that times should be entered in 24-hour format.
Example shows 9am to 5pm program
Set On and off times the same for
all unused time periods.
Figure 17 Daily Menu
On Time 1
09:00
Off Time 1
17:00
On Time 2
00:00
Off Time 2
00:00
On Time 3
00:00
Off Time 3
00:00
On Time 4
00:00
Off Time 4
00:00
On Time 5
00:00
Off Time 5
00:00
On Time 6
00:00
Off Time 6
00:00
3.3.2.1.4
Weekly Menu
Set the sampling days
Example shows Monday to Friday
program, with no sampling on
Saturday or Sunday.
Sunday
NO
Monday
YES
Tuesday
YES
Wednesday
YES
Thursday
YES
Friday
YES
Saturday
NO
Figure 18 Weekly Menu
3.3.2.2
Wind Menu
Note:
It is necessary to disable the Trigger function if
using the wind sensor to control the MicroVol. This is
achieved by setting the Trigger Lo and Hi parameters
to the same value, normally set them both to zero.
The Wind Menu contains wind sector and speed conditional sampling
parameters. When the wind direction angle has been greater than the
counter-clockwise wind direction, but less than the clockwise wind
direction, and has been greater than the Low wind speed limit but less than
the High wind speed limit, for between 60 and 80 seconds, the Wind
condition is satisfied. If all other conditions are satisfied (eg. programmable
26