17.  On/Off  : 

 

After turning on the device, the display shows the selected time and the unit is ready 
to work. Press the key 

 

twice

 and the cleaning will begin. Cleaning will stop when 

the time arrives. To stop the device before time runs out, press the key 

 twice

 

18.  Quick time selection key 

Press  the  key 

.  The  display  will  score 

.  Each  time  you  press  the  key,  it  will  increase  by  5 

minutes. 

 
 

OPERATION 

 

 

 

 
 
 

Remove the lid and hold it in the support. put the items in the basket. Then 
put  them  in  the  cleaning  tank.  Add  water  or  a  cleaning  solution  to  a  level 
between 

min.

  and 

max.

  and  above  the  area  to  be  cleaned.  In  order  to 

prevent scratches and friction of the products, do not pile them up 
If  the  device  runs  out  of  water,  the  energy  of  the  ultrasound  will  not  be 
absorbed. This model has a lot of power. In 15 seconds the device can be 
damaged or its life can be severely reduced. 

 

 

 

 

 

Connect  the  power  cord  and  turn  the  switch  on  (I).  The  display  will  show 

,  This  is  the  most  commonly  used  time.  If  the  time  needs  to  be 

adjusted, press If the time needs to be adjusted, press Time button and 

 

buttons 

 and 

. The time can be set between 1 to 30 minutes. Common 

cleaning time are 5, 10 or 15 minutes.  Prolonged cleaning time may result 
in: 

d.  Loosening of the screws if used. 
e.  Increasing the pre-existing cracks. 
f.  Peeling of coating which was already separated. 

 

 

 

Press 

 

button twice to start cleaning. 

During  cleaning,  buzzing  sound  can  be  Heard  from  the  cleaning  tank, 
indicating the cleaning is underway. Closing the lid will reduce the noise. 
LED  display  will  count  down to show the remaining  cleaning time. When it 
displays 

, the cleaning is done. To stop cleaning at any time, press 

 

button

 

twice. 

The unit is designed with overheating protection to avoid damage caused by 

extended  and  continuous  operation. When  the  light 

  is  on,  the  unit  will 

not turn on even it is forced to. The unit will idle for about 20 minutes. Then 
can be restarted after the green light comes on. 

Summary of Contents for R-100136

Page 1: ...DE LIMPIEZA POR ULTRASONIDOS 3 EN ULTRASONIC CLEANER 8 MESTRA TALLERES MESTRAITUA S L Txori Erri Etorbidea 60 Tel 34944530388 Fax 34 944711725 mestra mestra es www mestra es 48150 SONDIKA BILBAO ESPA...

Page 2: ...o drenaje Drainage pipe fitting E V lvula drenaje Drainage valve handle F Panel de control Control panel G Asas Handles H Interruptor general Power switch I Toma de corriente Power socket J Soporte ta...

Page 3: ...cuentra listo para funcionar PANEL DE CONTROL 1 Tiempo de trabajo Cuenta atr s desde que se inicia la limpieza 2 Term metro multiestado Verde Temperatura del agua entre 0 y 30 de la temperatura de con...

Page 4: ...vida til Conecte el cable y encienda el interruptor I El display mostrar que es el tiempo m s com n Si necesita ajustar el tiempo pulse las teclas y El tiempo puede ajustarse entre 1 y 30 minutos Tiem...

Page 5: ...agua sucia Limpie la cuba y s quela Cierre la v lvula de drenaje MANTENIMIENTO Y LIMPIEZA Para asegurar un ptimo rendimiento y una m xima fiabilidad del aparato le recomendamos No encienda el aparato...

Page 6: ...do a la red La tapa del aparato pude encontrarse caliente despu s de un ciclo Por este motivo ret rela con precauci n Cuando no conozca exactamente los efectos de la soluci n limpiadora sobre el objet...

Page 7: ...mpacto desintegra la suciedad en gr nulos y la separa de la superficie del objeto Los gr nulos pasan al agua CARACTER STICAS T CNICAS Tensi n de alimentaci n 230 V 50 Hz Potencia 300 W Frecuencia 35 k...

Page 8: ...Timer The timer counts down since the cleaning starts 11 State thermometer Green Water temperature between 0 and 30 of the set temperature Orange Water temperature between 30 and 60 of the set temper...

Page 9: ...Connect the power cord and turn the switch on I The display will show This is the most commonly used time If the time needs to be adjusted press If the time needs to be adjusted press Time button and...

Page 10: ...d maximum reliability of the device we recommend Do not turn the unit without water in the tank Even though the unit is designed with multiple protections if it is turned on for over 30 seconds withou...

Page 11: ...a cycle Proceed with caution when removing it If you are not sure of the effects of the cleaning solution on a given type of item carry out a previous test with a single item Do not attempt to remove...

Page 12: ...The impact dislodges the debris to achieve the cleaning effect TECHNICAL SPECIFICATIONS Power supply 230 V 50 Hz Power 300 W Frequency 35 kHz Outside dimensions 515 x 305 x 240 mm Tank dimensions 323...

Reviews: