96
Irinos IR Users Manual © 2015 - 2016 Messtechnik Sachs GmbH, Germany
Mounting & Cabling
benefits:
o
Cabling of the probes, encoders or any other sensors can be kept
short. This results in a higher quality of the analogue signals.
Further it helps to keep cables away from possible sources of
noise.
o
Changing a probe is easier, e.g. in case it is broken.
For a troublefree operation of the Irinos-System, please note:
Always keep a distance to possible sources of noise, e.g.
frequency changers or motor cables.
An important factor regarding the location selection, is the IP
protection class. Even though most of the Irinos components are
available with a high protection class, it is in reality difficult to achieve
it for the whole system. In order to comply with high IP protection
classes, it is important to use appropriate cables and connectors.
Most of the standard probes and incremental encoders are equipped
with a connector providing a low protection (e.g. IP00 or IP40).
These connectors would need to be replaced, which typically is not a
favourite choice for the end user.
Hence it is adviced to place the Irinos-Boxes at places where a low
or no IP protection is required.
The Irinos-Boxes have a low internal heat generation and are
designed for typical industrial ambient temperatures. Further the
internal measurement electronics has a high temperature stability.
Nevertheless a location with a moderate ambient temperature
should be chosen (see also the specifications in the datasheets). It
should particularly be avoided to place the Irinos components next
to sources of heat, e.g. heat sinks of other devices or heating
elements.
6.3
Mounting the Irinos-Boxes
6.3.1
Direct mounting via tapped bushes on rear side
Each Irinos-Box has 2 tapped busches M4 on the rear side. These
allow direct mounting.
Please observe the following limitations:
Summary of Contents for Irinos IR
Page 2: ......
Page 7: ...Introduction...
Page 15: ...Safety instructions...
Page 22: ......
Page 23: ...System overview...
Page 39: ...Product descriptions...
Page 64: ......
Page 65: ...Pin assignment...
Page 92: ......
Page 93: ...Mounting Cabling...
Page 109: ...Setup...
Page 118: ...118 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Setup...
Page 120: ......
Page 121: ...Measurement Control via MscDll...
Page 168: ......
Page 169: ...Troubleshooting and first aid...
Page 187: ...Maintenance cleaning and disposal...
Page 190: ......
Page 191: ...Application notes...
Page 198: ......
Page 199: ...Specifications dimensions...
Page 209: ...209 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...
Page 210: ...210 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...
Page 211: ...211 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...
Page 212: ...212 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...