112
Irinos IR Users Manual © 2015 - 2016 Messtechnik Sachs GmbH, Germany
Setup
Enumeration process
The duration of the enumeration process depends on the number of
Irinos-Boxes connected. Typically it takes only a few seconds. If
many Irinos-Boxes are used, it may last up to 15s.
Prior to being enumerated, the elements of the
turn clockwise:
Afterwards the Box number is shown in the 7-digit display. An
exception is the first Box having a 7-digit display: it first shows the
IP configuration. Depending on the configuration, this lasts a few
seconds.
Termination
During the enumeration process, the first and the last Irinos-Box of
the ILink cabling are automatically detected and terminated. This is
signalled by a blue LED between both ILink connectors at these
Boxes. The LED must be off at all other Boxes.
Checks
The following checks should be done after startup:
o
All slave Boxes must display a valid address.
o
The termination LED must be turned on at the first and the last
Irinos-Box of the ILink cabling.
7.3
Network configuration
The Master-Box of the Irinos-System provides an integrated DHCP-
Server, which is activated at delivery state. The
IP address of the
Irinos-Systems is 192.168.3.99
, the subnet mask is
255.255.255.0. If the Ethernet-Interface of the PC is configured as
"DHCP-Client", a valid IP address is assigned to it (range
192.168.3.100 to 192.168.3.254). No further network
configuration is required.
In delivery condition the DHCP-Server of the Irinos-System
is activated. This is a good choice for a direct connection to
a PC.
42
Summary of Contents for Irinos IR
Page 2: ......
Page 7: ...Introduction...
Page 15: ...Safety instructions...
Page 22: ......
Page 23: ...System overview...
Page 39: ...Product descriptions...
Page 64: ......
Page 65: ...Pin assignment...
Page 92: ......
Page 93: ...Mounting Cabling...
Page 109: ...Setup...
Page 118: ...118 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Setup...
Page 120: ......
Page 121: ...Measurement Control via MscDll...
Page 168: ......
Page 169: ...Troubleshooting and first aid...
Page 187: ...Maintenance cleaning and disposal...
Page 190: ......
Page 191: ...Application notes...
Page 198: ......
Page 199: ...Specifications dimensions...
Page 209: ...209 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...
Page 210: ...210 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...
Page 211: ...211 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...
Page 212: ...212 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...