Version 2.1
en MBI 46.31 operating instruction.docx
Issue date: 01/2022
Page 20 of 42
Terminal
Assignment
17
PE conductor
18
230 VAC
19
230 VAC
22
-15 V
23
+15 V
24
Ground GND
Figure 11
Pin assignment of 230 V power supply
Amplifier
Linear Variable
Differential
Transformers
all Dxx
Linear Variable
Inductance
Transducer
all Wxx except WP
Long-stroke
sensor e.g.
WP
Plug Terminal Wire
strand
Cable Plug
Wire
strand
Cable Plug Cable Plug
ac 2
16
WH
(BN)
WH
2 (B)
BU
BU
3 (C)
BU
C
ac 4
15
BU
BU
3 (C)
RD
RD
2 (B)
RD
B
ac 6
14
RD
RD
1 (A)
-
-
-
-
-
ac 8
13
BK
BK
4 (D) WH+YE WH
1 (A)
WH
A
Figure 12
Connection of MESSOTRON displacement sensors
5.2 Setting options of the measuring amplifier
The measuring amplifier must be adapted to the (displacement) sensor used.
The following parameters must be set:
•
Position of the electrical zero point,
•
Phase compensation (possibly caused by the sensor type and the
connecting cable),
•
Gain (compensation of sensor sensitivity).
At the factory, a
basic configuration
for all three parameters is set in the
measuring amplifier. This is achieved by using a variable component assembly
(resistors / capacitors) based on:
•
a reference displacement sensor, if the sensor type is known,
•
the sensor specification provided by the customer,