Spectroquant
®
Colorimeter Picco COD / CSB
4
Special function adjustment mode
4.1 User
adjustment
4.1.2 Checking / changing the adjustment mode
Resetting the user adjustment to the original factory adjustment will reset all
methods and ranges.
A user-adjusted method is indicated by an arrow while the test result is dis-
played at the position “Cal”.
•
User adjustment is active:
•
Factory adjustment is active:
The instrument can be reset to the delivery status (factory adjustment
setting) in the following way:
After the instrument has been switched off,
hold
the [SELECT] and
[ZERO/TEST] keys
depressed
.
Switch the instrument on using the [ON/OFF] key, after approx. 1 second
release the [SELECT] and [ZERO/TEST] keys.
The display shows in alternating mode:
the instrument is in delivery status (“SEL” stand for “select”)
or
the instrument is operating with a user-specified adjustment.
If this is to be retained, switch the instrument off with the [ON/OFF] key.
Adjustment is reset to the factory setting for all methods at the same time by
pressing the [SELECT] key.
The display shows in alternating mode:
Switch the instrument off with the [ON/OFF] key.
CAL
SEL
(3 sec.)
cAL
SEL
CAL
SEL
Select
Select
Zero
Test
On
Off
On
Off
On
Off
On
Off
Select
RESULT
RESULT
Cal
Cal
CAL
cAL
Summary of Contents for Spectroquant Picco COD/CSB
Page 2: ......
Page 18: ...Ausgabedatum 03 2009 18 Spectroquant Colorimeter Picco COD CSB de...
Page 34: ...Release 03 2009 34 en Spectroquant Colorimeter Picco COD CSB...
Page 50: ...Version du 03 2009 50 Spectroquant Colorim tre Picco COD CSB fr...
Page 66: ...Edizione 03 2009 66 Spectroquant Colorimetro Picco COD CSB it...
Page 82: ...Fecha de emisi n 03 2009 82 Spectroquant Color metro Picco COD CSB sp...