Operator’s manual

 

Multitest plus

 

 

Atemschutztechnik MENZL GmbH 
Absbergerstraße 9 
3462 Absdorf  
AUSTRIA

 

 

 

Tel.: 

+43 (0) 2278 / 2102-0 

 

Fax: 

+43 (0) 2278 / 2102-21 

 

E-mail: 

[email protected]

 

 

Web: 

 

www.menzl.at/en

 

Summary of Contents for 330-057

Page 1: ...Our expertise your safety since 1975 Operator s manual Multitest plus Atemschutztechnik MENZL GmbH Absbergerstra e 9 3462 Absdorf AUSTRIA Tel 43 0 2278 2102 0 Fax 43 0 2278 2102 21 E mail info menzl...

Page 2: ......

Page 3: ...governed demand valve without negative medium pressure 9 Lung governed demand valve leak tightness with medium pressure 10 Lung governed demand valve opening pressure 10 Medium pressure test 11 Mask...

Page 4: ......

Page 5: ...Whenever it is likely that the tester is no longer safe for use make it inoperative and secure it against any unintentional operation A service representative of the manufacturer or a similarly traine...

Page 6: ...uge Pressure range Accuracy class Low pressure 20 mbar 1 6 Medium pressure 0 16 bar 1 Scope of delivery The scope of delivery includes Multitest plus accessories for testing may vary depending on the...

Page 7: ...tment on the side of the tester 4 pressure pump lever 9 jack for power adapter 1 5V 5 relief valve 10 medium pressure connector for testing hose 20 dummy head 30 mask testing hose 21 push button valve...

Page 8: ...her types of connectors as well as different makes of chemical protective suits Features of the tester One battery type D cell is required for operating the tester A compressed air supply is not neede...

Page 9: ...quipment connector Seal the connecting nipple of the lung governed demand valve Move the pressure pump lever to the OPEN position Press and hold the pump push button Let the pressure build up until it...

Page 10: ...the pressure pump lever to the CLOSED position Release the pump push button Release pressure via the relief valve until the level required for testing is reached Perform the test against the specifica...

Page 11: ...governed demand valve Perform the test against the specifications provided by the equipment manufacturer Mask leak tightness Strap the mask onto the dummy head Connect the testing hose to the tester a...

Page 12: ...m pressure IMPORTANT Be sure to strictly adhere to the sequence of steps Detach the lung governed demand valve from the equipment connector Deactivate the positive pressure function of the lung govern...

Page 13: ...valve to the equipment connector Release pressure via the relief valve for 5 seconds Perform the test against the specifications provided by the equipment manufacturer Medium pressure test Do NOT atta...

Page 14: ...by the equipment manufacturer Mask exhalation valve opening pressure Strap the mask onto the dummy head Connect the testing hose to the tester and the mask Move the pressure pump lever to the OPEN pos...

Page 15: ...ective suit valve testing Connect the testing hose to the inside part of the valve Attach the testing hose to the tester s equipment connector Move the suction pump lever to the OPEN position Press an...

Page 16: ...dummy head by pressing the push button The dummy head is insufficiently inflated The dummy head is ideally shaped for testing masks The dummy head is inflated too much CAUTION Inflating the dummy hea...

Page 17: ...it Replace the battery Remove the empty battery and insert a new one Always observe the correct polarity Put the battery compartment cover back on Put the battery compartment cover back on and turn it...

Page 18: ...Interspiro valve testing adapter 330 502 Equipment for testing chemical protective suits CPS Testing hose with connecting adapter for Auer CPSs 330 039 Filling hose with connecting adapter for Auer CP...

Page 19: ...tic operation Disinfectant gets added automatically Freely adjustable concentration level 0 5 to 5 Masks are dried at a maximum of 50 C by means of a hot air blower 330 487 Detergent EW80 clean For cl...

Page 20: ...ung governed demand valves LGDV pressurized during cleaning for 3 LGDVs 330 535 for 6 LGDVs 330 536 for 9 LGDVs 330 462 Sealer For sealing respiratory protective masks lung governed demand valves and...