mensor CPG2300 Operating Instructions Manual Download Page 1

Operating Instructions

Portable Digital Pressure Gauge

Portable Digital Pressure Gauge Series 2300 

PN 0017441003N • 01/2017

CPG2300

Summary of Contents for CPG2300

Page 1: ...Operating Instructions Portable Digital Pressure Gauge Portable Digital Pressure Gauge Series 2300 PN 0017441003N 01 2017 CPG2300 ...

Page 2: ... considerable damage mortal danger danger of injury will occur if the respective safety precautions are not taken Caution This Caution symbol indicates danger for the system and material if the respective safety precautions are not taken This Notice symbol does not indicate safety notices but information for a better understanding of the facts Copyright 2005 Mensor All rights reserved ...

Page 3: ... Warnings and Caution Notices 7 3 Product Description 9 3 1 General Description 9 3 2 Display 9 3 3 Power Supply 9 3 4 Battery 9 3 5 Serial In Out Communications 10 4 Specifications 11 5 Installation 13 5 1 Unpacking the Instrument 13 5 2 Dimensions 14 5 3 Pressure Connections 15 5 4 Electrical Connections 16 5 5 Power On 16 6 Local Operation 17 6 1 Keypad 17 6 2 Standard Display 18 6 3 Menu Scree...

Page 4: ...26 7 2 3 Commands and Queries 27 7 2 4 User Programmable Pressure Units List 30 7 2 5 Pressure Unit Codes and Conversions 31 8 Options 33 8 1 Current Input 33 8 1 1 Menu Screens 33 8 1 2 Operational Notes 35 8 1 3 Connection Configurations 36 8 1 4 Application Diagrams 37 9 Calibration 39 9 1 Environment 39 9 2 Equipment 39 9 3 Pressure Standard 41 9 4 Calibration Medium 41 9 5 Calibration Process...

Page 5: ...he product the cause and nature of a defect and the necessity or manner of repair Service repairs or disassembly of the product in any manner performed without specific factory permission voids this warranty MENSOR MAKES NO WARRANTY OF ANY KIND WITH REGARD TO THIS MANUAL INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE Mensor shall not be ...

Page 6: ...siness profits business interruption loss of business information and the like arising from or in connection with the delivery use or performance of the software program 1 4 Mensor Service Plus If you have problems and you don t find the answer in this manual contact Mensor at 1 800 984 4200 USA only or 1 512 396 4200 for personal assistance or at any of the contact addresses listed on the rear co...

Page 7: ...r Externally mounted relief valves to provide overpressure protection are avail able from Mensor as optional devices and are highly recommended for very low pressure sensors less than 10 psi Caution CAUTION It is recommended that a backup wrench be used when installing fittings or connections on the CPG2300 Caution CAUTION Use of a power supply other than the one provided by Mensor will invalidate...

Page 8: ...8 Operating Instructions CPG2300 Series Portable Digital Pressure Gauge CPG2300 Series Notes ...

Page 9: ... from 5 to 1000 psig 7 5 to 1000 psia Additionally the CPG2300 can be equipped with optional current measurement from 4 to 20 mA and a 24V dc loop power See the Options section for more information 3 2 Display Several different display screens are presented while operating the CPG2300 Each screen is arranged into blocks of information generally organized into columns and rows These screens are eac...

Page 10: ...tage vs Temperature The battery can be charged over an ambient tempera ture range of 0 to 40 degrees C with a supply voltage of 9 to 12 Vdc 3 5 Serial In Out Communications The serial port is configured as RS 232 9600 baud 8 data bits 1 stop bit and no parity It is designed as a 9 pin DCE device so that a straight pin to pin cable can be used to con nect it to a standard PC serial port The serial ...

Page 11: ... 1 psig over the compensated temperature range for 180 days Pressure Ranges Channel A 7 5 to 6015 psia 0 36 to 6000 psig Channel B 7 5 to 1000 psia 5 to 1000 psig Calibration Interval 180 days Resolution 5 digits on display 6 digits over serial port Overpressure Limit 1 5 x span Storage 20 70ºC Warmup 1 minute Reading Rate 4 6 second Response Time 252 mS Orientation Negligible 30 psi zero offset r...

Page 12: ...One year Options Carrying pouch attaches to belt Hand pump Relief valves mounted externally Current Input Option Battery life 4 hours with loop power 20mA and 10 hours with loop power 4mA Current measurement 4 to 20 mA calibrated Loop power voltage 24V unregulated Current uncertainty 0 03 of R Units psi inHg 0C and 60F inH2O 4C 20C bar mbar mmHg 0C cmHg 0C hPa kPa Pa MPa kg sq cm cmH2O 4C cmH2O 20...

Page 13: ...sting each unit is inspected for appearance prior to leaving the factory Upon receipt please examine the transducer for shipping damage Report any ap parent damage to the carrier immediately In addition to this manual you should have CPG2300 Digital Pressure Gauge Power adapter charger Any accessories ordered An envelope containing the Calibration Certificate ...

Page 14: ...14 Operating Instructions CPG2300 Series Portable Digital Pressure Gauge CPG2300 Series 5 2 Dimensions Figure 5 2 Case Outline Dimensions ...

Page 15: ...ave a second 1 4 inch FNPT threaded port on the right to measure the pressure on Channel B If its range is less than 20 psi and it is a gauge sensor a barb fitting provides access to the reference port of the sensor Caution CAUTION Avoid excessive overpressure to the sensor Externally mounted relief valves to provide overpressure protection are avail able from Mensor as optional devices and are hi...

Page 16: ...onnected to the CPG2300 RECEIVE line TX to RX and vice versa The power connector is a 2 1 mm standard power jack The center pin is 12 Vdc and the outside is ground see Figure 5 4b Figure 5 4a RS 232 Hookup Figure 5 4b DC Input 5 5 Power On After the pressure connections are secure press the power switch The CPG2300 will run through a brief initialization sequence and then will display the sensor c...

Page 17: ...e navigation in the menus They are also used to adjust the displayed pressure for calibra tion adjustment and adjustment of the programmable user unit The Main Menu Key The M key beings up the main menu The Enter Key The central key is used as an enter key to select the highlighted menu choice or for single button null Display Screen Displays user data and menu setup screens to access or change va...

Page 18: ...m highlight the item and press the enter key 6 3 1 Main Menu Screen If the M menu key is pressed at any time the main menu is displayed on the screen If one sensor is installed in the CPG2300 the Sensor A setup screen is displayed If two sensors are installed in the CPG2300 the following will be displayed Main Menu Sensor A A Setup Sensor B B Setup Sensor A B The items in this menu perform the fol...

Page 19: ...sor A Setup Peak On Peak Rst Null Off Cal Units The current condition of the peak indicator is displayed If the menu item displays Peak Off the peak indicator is disabled for the active sensor If Peak Off is selected it will enable peak capture for the active sensor If the Sen sor Setup menu is re entered the menu item will display Peak On to indicate that the peak capture function is enabled Peak...

Page 20: ...ing a null offset Single button null The null feature can also be accessed by pressing the enter key while at the standard Display When both channels are displayed pressing the null button once will null channel A Pressing the null button a second time will null channel B Pressing the null button a third time will un null channel A and a fourth press will un null channel B When a single channel is...

Page 21: ...sensor to be adjust ed the calibration to be reset to the factory defaults and the calibration password to be changed A Calibration Cal Zero Cal Span Change PW Reset Cal The items in the calibration menu perform the following functions when selected Cal Zero Displays the measured pressure for the active sensor Adjust the digits to the true pressure applied to the active sensor and press the enter ...

Page 22: ...nu is then displayed Caution The password is set to 0000 at the factory When making a change write it down and save the number If the password is lost contact Mensor Reset Cal Displays a confirmation screen A Calibration Cal Zero Cal Span Change PW Reset Cal If the enter key is pressed the calibration is reset to the factory defaults and the main operation screen is displayed If any other key is p...

Page 23: ...election and entry of a special User unit The User unit has a config urable conversion factor that converts from PSI to the desired value To enter the conver sion factor select Set User on the last units screen to display Set User Units A 1 00000e 1 Enter to Continue Use the left and right arrow keys to move the cursor under the digit to change and the up and down arrow keys to increment the digit...

Page 24: ...24 Operating Instructions CPG2300 Series Portable Digital Pressure Gauge CPG2300 Series 6 4 Menu Tree Figure 6 4 CPG2300 Menu Tree CPG2300 ...

Page 25: ...significant to the CPG2300 All other punctuation charac ters are ignored Some commands require a password string A carriage return cr and or linefeed lf must terminate each command or query The global address character is an asterisk It can be sent in place of the actual CPG2300 address All of the commands are channel dependent If sensor A is selected by sending XCH A all subsequent queries and co...

Page 26: ... to indicate that it has received a valid command or password rather than a query If an invalid command or password is sent the R character will not be returned When communicating with the CPG2300 the R character must be received before sending additional commands or queries In the event that an invalid command could be sent a time out period of approximately one sec ond should be used before send...

Page 27: ...f the LCD display is displaying either sensor A or B only and the DSP query returns 1 or 2 the return string will return a single value from the selected sensor Example format 1 14 4711 cr lf If the peak function is enabled the return string will contain three values The first value is the maximum pressure the second is the present pressure and the third is the mini mum pressure This return string...

Page 28: ...p DC sp nn nnnn cr lf Returns the date of calibration where nnnnnn is mmddyy Yes DC XDC sp nnnnnn cr lf R Loads the date of calibration where nnnnnn is mmddyy No DSP XDSP cr lf X sp DSP sp N cr lf Returns display mode where 1 Sensor A 2 Sensor B 3 Sensors A B No DSP XDSP sp n cr lf R Loads display mode where 1 Sensor A 2 Sensor B 3 Sensors A B No FL XFL cr lf X sp FL sp nn cr lf Returns the filter...

Page 29: ...for the active sensor A space must be included after the RST command No SAVE XSAVE cr lf R Saves all data to memory No SC XSC cr lf X sp SC sp nnn nnnn cr lf Returns the span correction multiplier for the active sensor Yes SC XSC sp nnnnnn cr lf R Sets the span correction multiplier for the active sensor Yes SCPW XSCPW sp nnnn cr lf R Sets the customer password No T XT cr lf X sp T sp n cr lf Retu...

Page 30: ...er can select from one to eight units to be displayed by sending a sequence of UT commands The syntax is as follows UT sp n sp n cr lf where the first n is a value from 0 to 7 and represents the units sequential position in the displayed list The second n represents the unit code from Table 7 2 5 If the first 0 unit in the sequence is assigned a value of 0 then the displayed units will default to ...

Page 31: ...e Unit PSI Conversion Factor 0 Counts N A 1 psi 0 2 inHg 0C 2 036020 3 inHg 60F 2 041772 4 inH2O 4C 27 680672 5 inH2O 20C 27 729767 14 bar 0 06894757 15 mbar 68 94757 17 cmH2O 4C 70 3089 19 mmHg 0C 51 71508 20 cmHg 0C 5 171508 22 kPa 6 894757 23 Pa 6894 757 26 kg sq cm 0 07030697 27 mSW 0C 3 5 salinity 0 6838528 34 hPa 68 94757 36 MPa 0 006894757 38 cmH2O 20C 70 4336 99 User User settable ...

Page 32: ...32 Operating Instructions CPG2300 Series Portable Digital Pressure Gauge CPG2300 Series Notes ...

Page 33: ...nput section is protected from over current or high voltage input by a comparator controlled mosfet that will limit the current to a nominal of 24 8 mA When the resident loop power is enabled the current limiter circuit will protect the instru ment if the input terminals are shorted together 8 1 1 Menu Screens Main Menu The current input is set up using the mA Input selection in the main menu This...

Page 34: ... entry screen mA Input Setup Display mA Off Source V Off Calibrate mA Password Entry Screen If the calibrate mA function is selected from the mA Input Setup screen the password entry screen will be displayed mA Calibration Enter to Continue mA Calibration Screen The mA calibration screen allows zero and span to be ad justed the calibration to be reset to the factory defaults or the calibration pas...

Page 35: ...the alternate channel to the pressure channel that is selected This means when Sensor A is selected as the pressure measurement channel the pressure will be displayed as the top reading and the current measurement will be the bottom reading The display order is reversed when Sensor B is selected Both pressure channels can also be displayed while the current measurement channel is active by selecti...

Page 36: ...ack red is inserted on the high side and the Sense jack black is inserted on the low side in the loop The pressure port fittings and the d sub shell of the RS 232 connector are electrically common to the Sense connection each through 1 Meg ohm resistors The RS 232 digital ground is connected directly to the Sense The connections should be isolated from the current loop connections Figure 8 1 3 Ele...

Page 37: ...ating Instructions CPG2300 Series 37 Portable Digital Pressure Gauge CPG2300 Series 8 1 4 Application Diagrams Figure 8 1 4a Connections in a powered loop Figure 8 1 4b Connections using CPG2300 loop power ...

Page 38: ...38 Operating Instructions CPG2300 Series Portable Digital Pressure Gauge CPG2300 Series Notes ...

Page 39: ...period may be extended as confidence is gained in the span stability 9 1 Environment Whenever possible calibrate the CPG2300 at the same ambient temperature as its normal operating environment The temperature should be stable and within the specified calibration range In addition for maximum accuracy see that the CPG2300 is at rest on a stable platform which is free of vibration and shock and orie...

Page 40: ...40 Operating Instructions CPG2300 Series Portable Digital Pressure Gauge CPG2300 Series Figure 9 2 Calibration Setup MODEL 2300 PRESSURE GAUGE MODEL 2300 PRESSURE GAUGE ...

Page 41: ...ill be displayed Enter the four digit password and then press the enter key For more information on entering or changing the password see the Local Operation section 9 5 1 Zero Adjustment Access the zero calibration screen as described in the Local Operation section For gauge sensors Vent the CPG2300 and adjust the display to show zero pressure by using the arrow keys on the front panel When zero ...

Page 42: ...0 984 4200 Fax 512 396 1820 sales mensor com www mensor com Operating Instructions CPG2300 Series PN 0017441003N 01 2017 WIKA Alexander Wiegand SE Co KG Alexander Wiegand Straße 30 63911 Klingenberg Germany Tel 49 93 72 132 5015 Fax 49 93 72 132 8767 CTSales wika de www wika de ...