Operating Instructions - CPC3050
23
High-Speed Pressure Controller
CPC3050
5.3 Mounting
The instrument can be set up on a desk top or it can be rack-mounted. Rack mount hardware is optional
on the CPC3050 (see Section 5.2 Dimensions and Section 8, Options).
The special transducers used in the CPC3050 are relatively insensitive to tilt and vibration. However to
further assure stability and accuracy, avoid mounting the instrument on surfaces subject to excessive mo-
tor or machinery vibration.
5.4 Rear Panel
Five pneumatic pressure ports are located across the rear panel. Any instrument containing sub-atmo-
spheric pressure ranges needs a vacuum pump to be connected at the exhaust port. The instruments
with pressures above or at atmospheric pressures can have the exhaust port open. Next to the ventilation
fan is a 10-32 UNF fitting which is connected to the barometric reference if installed. Below the ventilation
fan is the connection to drive the automatic contamination prevention system. Positioned on the right side
is a remote communication connections; RS-232, Ethernet, IEEE- 488, USB along with the USB host
connection.
Figure 5.4 A – Rear Panel
5.4.1 Pressure Connections
Warning
!
Warning! The pressure connections must be installed according to the following
instructions, observing the relevant regulations. The installation is to be per-
formed by trained, authorized personnel, knowledgeable in the safety regulations
for working on pneumatic/hydraulic systems.
Check for trapped pressure when loosening any pressure connection.
i
Notice
Five pressure connections are on the rear panel.
All pressure ports on the rear apart from the Vent Port are female 7/16 - 20 SAE/MS straight threads per
MS16142 and SAE J514 table 14. Connected adaptors require a tube fitting boss seal with an o-ring per
MS33656. Mensor can provide a variety of adapter fittings (see Section 8 Options) with the instrument.
Do not use sealant on fittings sealed with an o-ring. The integrity of each seal is particularly important
since even microscopic leaks can cause errors in pressure measurements.