Menu functions
7.3.5
Timer mode
By default, the timer only starts when the tolerance band of ±3 K
around the setpoint temperature has been reached (Fig. 26). To
ensure that the required temperature is maintained for a
sufficient period of time, this setting cannot be changed. If the
temperature is no longer within the tolerance band, the
sterilisation time is, for reasons of safety, restarted as soon as
the required temperature is reached again.
t
Timer
°C/°F
t
6 K
T
set
Fig. 26
Timer mode: The timer only starts when the tolerance band of ±3 K around the set
temperature has been reached
7.3.6
Type of the slide-in unit (
Grid
or
Shelf
)
Here, you have to set the type of the slide-in unit (grid or shelf)
used. The selection Shelf enables you to adjust the control
function to the different air flow characteristics in the interior
when using optional sliding shelves instead of the grids that are
part of the standard delivery.