6.  Hints about fault corrections 

6.1.  Timer does not automatically switch off power ________________________   

If the timer is defective, the sterilizer can still be 

operated in a half-automatic way with the help of the 

permanent switch (6). Before exchange of the timer, 

it is absolutely necessary to pull out the main plug. 

Loosen the screw of the timer knob and remove the 

knob. Unscrew the front scale plate and remove it 

with care. Then remove the stick hulls of the cable 

ends from the flat plugs of the timer and unscrew the 

timer from the front scale plate. Connect and install 

the new timer in the same way. 
A new timer can only be ordered at specialist shops. 

For quick completion, it is necessary to know the 

type of sterilizer and its work number. 

6.2.  Temperature set up with the regulation knob does not tally with 

indication of the thermometer _______________________________________   

Hot-air sterilizers from MELAG are delivered with 

temperature calibrated at 180°C by adjusting the 

regulator axis so that the thermometer really 

shows a value of 180°C. Then the thermostat 

knob is sticked on the axis so that the white 

indicator of the knob points to the 180°C- mark on 

the scale. In this position, the thermostat knob is 

fixed to its axis by fastening the clamp screw. 
If the thermostat knob is unintentionally be turned 

with force beyond the resistance due to improper 

treatment, e.g. during cleaning or unpacking of the 

gadget, or if the screw is loosened in the taper, the 

knob can slip on the regulator axis. Respectively, 

the real temperature inside the vessel (indicated 

on the thermometer) is different from the one set 

up on the regulator scale. 
A quick correction can easily be achieved by 

turning the thermostat knob either little clockwise 

in order to obtain a higher temperature, or 

counter-clockwise for lower temperature. Provi-

sional markings on the scale will help adjustment. 
A correct calibration is carried out as follows: 
Remove the grey cap from the thermostat knob. 

After loosening the clamp screw below, the knob 

can be removed from the axis. Close the door and 

switch on the sterilizer. Now adjust the tem-

perature regulator by turning the axis (perhaps 

with small pliers) so that the thermometer shows 

the desired temperature (e.g. 180°C) after con-

clusion of pre-heat and equalization time. Put the 

thermostat knob back on the axis so that the white 

indicator points to a value  on the scale which 

agrees with the temperature shown on the 

thermometer. Finally screw the knob's screw tight 

without changing the position of the axis. 

6.3.  The appliance does not heat up _____________________________________   

If your hot air steriliser does not work, check the 

internal fuses. To change fuses, first disconnect 

the appliance from the mains power supply. Press 

down the tongue on the fuse container on the rear 

side of the oppliance and pull the container out. 

Inspect the fuses and exchange if necessary. 

If the signal lamp (4) does not light up to show that 

the power is on, check the cable, power socket, 

and mains fuse. 
If the signal lamp (4) lights up but not signal lamp 

(2), check the setting of the control knob (see 1. 

above). 

7.  Appendix 

7.1.  Spare parts ______________________________________________________   

Article No. 

Description 

32985 

door seal (type 215-225) 

92995 

door seal (type 255) 

32890 

thermostat 

32930 

timer  

33470 

thermometer 

16450 

mains fuse 6,3A 2x 

 