background image

2 General guidelines

6

Symbols on the device

Manufacturer of the product

Date of manufacture of the product

Product serial number from the manufacturer

Article number of the product

The user manual includes important safety information. Failure to comply with these
instructions can result in injury and material damage.

Please read this user manual carefully before commissioning the device.

This symbol indicates that the marked area becomes hot during operation. Contact with
it during or shortly after operation can pose the danger of burns.

In affixing the CE mark, the manufacturer declares that this product fulfils the basic
requirements of the EMC and low voltage directive.

The device may not be disposed as domestic waste. The vendor is responsible for
appropriate disposal of the device - it must be delivered to the vendor to be disposed of.
In affixing this symbol, the manufacturer furthermore declares that he has satisfied all the
legal requirements pertaining to the release, redemption and environmentally sound
disposal of electric and electronic appliances.
MELAG devices are synonymous for long-term quality. When you eventually need to
decommission your MELAG device, we offer a special disposal service. Simply contact
your stockist.

Summary of Contents for Sterilizer 255

Page 1: ...shed MELAG as the world s leading manufacturer in the instrument treatment and hygiene field You our customer are justified in your demand for the best products quality and reliability Providing compe...

Page 2: ......

Page 3: ...air sterilizer on and off again 11 Opening and closing the door 12 Changing the mount 12 6 Sterilization 13 Preparing the sterilization material 13 Loading the dry heat sterilizer 13 Information regar...

Page 4: ...scope of delivery may be connected to the device n The power cable may not be replaced by a cable determined to be insufficient Danger of short circuit n Liquids may not be permitted to reach the inte...

Page 5: ...perating Manual near to the device It represents a component of the product Formatting rules Example Explanation see Chapter 2 Reference to another text section within this document Symbols used Symbo...

Page 6: ...the danger of burns In affixing the CE mark the manufacturer declares that this product fulfils the basic requirements of the EMC and low voltage directive The device may not be disposed as domestic...

Page 7: ...onally Tray Tray jack Intended use This hot air sterilizer is intended for use in the cosmetics sector the cosmetic chiropody and the veterinary sector It was designed for the sterilization of objects...

Page 8: ...lamp heating 3 Thermometer 4 Indicator lamp network 5 On Off switch and time setting 6 Door opens forwards 7 Door handle 8 11 9 10 Fig 2 Rear view of the device 8 Fan 9 Device foot 10 Type plate 11 Co...

Page 9: ...at least six trays are or dered with the device the scope of delivery will include a tray jack The following loading versions are possible in the device Standard mount The series equipment comprises...

Page 10: ...clearance of 10 cm to other devices and walls especially flammable parts u Ensure sufficient ventilation Ensure sufficient clearance to the top for free removal of the warm air The device may not be u...

Page 11: ...wise The control lights network and heating will remain illuminated until the time expires The hot air sterilizer can be set to continuous operation In this case the hot air sterilizer must be switche...

Page 12: ...cedure has started this could result in cooling and the object not being steri lized for a sufficient time 1 Move the On Off switch to the 0 position to switch off 2 Open the door forwards to open 3 M...

Page 13: ...ver place the sterilization material together in blocks as this will prevent heat equalization u Do not place the sterilization material on cellulose as this can result in a heat build up u Always pac...

Page 14: ...the On Off switch Comply with the information of the sterilization times table Selecting the sterilization times Do not exceed the specified loading quantities Use only trays and cartridges made of a...

Page 15: ...a specified below for the storage of sterilized equip ment u Comply with the maximum storage duration in accordance with the packaging type u Do not store the sterilized equipment in the decontaminati...

Page 16: ...ng bio indicators thermocouples or maximum thermometers Comply with the valid regional legal specifications When performing the spore test using biological indicators the spore packets are to be clamp...

Page 17: ...o remove the impurities with this method Only if the chamber mount or chamber seal face has persistent soiling should you use a mild stainless steel cleaning agent with a pH value between 5 and 8 To c...

Page 18: ...temperature has been set too low The temperature controller rotary knob was unintentionally turned or mal adjusted during unpacking or incorrect handling e g during cleaning Set a higher temperature T...

Page 19: ...ght 22 kg Please observe the maximum load 6 kg including 6 trays Maximum mains voltage supply variations 10 Maximum voltage range 207 253 V Relative humidity up to 31 C max 80 up to 40 C max 50 Degree...

Page 20: ...11 Accessories 20 11 Accessories Table 2 Accessories Designation of article art no Tray 02003 Tray jack 28890...

Page 21: ...e sterilization material to the required temperature of 180 C Sterilization material Unsterile sterilizable material which is still to be sterilized Sterilized equipment Also referred to as a batch a...

Page 22: ...0829 Berlin Germany Email info melag com Web www melag com Original instructions Responsible for content MELAG Medizintechnik GmbH Co KG We reserve the right to technical alterations BA_Hei luftsteril...

Reviews: