Chapter 2 – Installation 

4. 

Connect the waste water line (b

1

) and the 

cold water line (c

1

) to the underside of the 

steam sterilizer. 

5. 

Only at Cliniclave 45 M: Place the hoses 
over the bracket (d) in the rear area of the 
steam sterilizer, so that it protrudes from the 
floor unit. 

 

 

 

 

6. 

Lead the waste water line (b

2

) and the cold 

water line (c

2

) through the duct on the right-

hand side of the osmosis unit and connect it 
to the corresponding connections on the 
osmosis module. 

 

 

7. 

Connect the end of the feed water hose to 
the free screw connection (e) behind the 
manometer. 

 

WARNING 

The hoses may not be allowed to suffer 
kinking or crushing. 

 

 

 

8. 

Connect the power plug of the reverse 
osmosis unit with the power plug of the 
steam sterilizer (f) on the underside of the 
steam sterilizer and then connect the 
bracket. 

 

  DANGER 

When the steam sterilizer is switched on, 
a voltage of 230 V is found on the power 
plug of the reverse osmosis unit to the 
Cliniclave 45 M/Cliniclave 45 MD. 

 

 

 

 

b

1

 

c

1

 

b

2

 

c

2

 

 

Summary of Contents for MELAdem 56 M

Page 1: ...e instrument treatment and hygiene field You our customer are justified in your demand for the best products quality and reliability Providing competence in hygiene and Quality made in Germany we guar...

Page 2: ...smosis unit MELAdem 56 M About this manual Symbols used Symbol Explanation Draws your attention to a situation which if not avoided could result in damage to the reverse osmosis unit Draws your attent...

Page 3: ...8 Chapter 3 Commissioning 10 Rinsing the reverse osmosis unit 10 Checking the operational readiness 11 Chapter 4 Maintenance 12 Regular checks and maintenance 12 Replacing the fine filter pre filter...

Page 4: ...mosis unit please observe the following safety instructions as well as those contained in subsequent chapters Maintenance Use only original replacement parts for maintenance Leaks Close the water inta...

Page 5: ...s disposed of into the effluent system Mode of functioning An ion exchanger is connected to the reverse osmosis unit to achieve the level of water quality necessary for operation of the Cliniclave 45...

Page 6: ...Chapter 1 Device description 6 Views of the device Fig 1 Reverse osmosis unit MELAdem 56 M...

Page 7: ...mosis unit 8 Pre filter Protects the sensitive reverse osmosis membrane The pre filter filter holds back all particulate material rust and other soiling 9 Recessed grip Acts as a carry aid for transpo...

Page 8: ...ows to connect the reverse osmosis unit 1 Switch off the steam sterilizer at the power switch 2 Remove the caps from the hose connections on the underside of the steam sterilizer and the reverse osmos...

Page 9: ...unit and connect it to the corresponding connections on the osmosis module 7 Connect the end of the feed water hose to the free screw connection e behind the manometer WARNING The hoses may not be all...

Page 10: ...unit 4 Place and hold the free end of the feed water line in a 5 l container or similar 5 Switch on the steam sterilizer at the power switch 6 Operate the reverse osmosis unit for c 20 minutes in orde...

Page 11: ...disappears on the steam sterilizer display a program can be started The reverse osmosis unit now works fully automatically Checking the operational readiness Operating pressure Perform a daily check o...

Page 12: ...reverse osmosis unit can be found in the maintenance instructions of the steam sterilizer Poor local water quality requires bi annual maintenance of the reverse osmosis unit in accordance with sectio...

Page 13: ...Switch on the steam sterilizer and open the shut off valve at the manometer 9 Check whether all the parts are secure and that the reverse osmosis unit is free from leaks Replacing the mixed bed resin...

Page 14: ...k the filter housing 3 and 5 for soiling and rinse out if necessary 7 Insert the filter insert 4 and sealing ring 2 Replace a worn sealing ring 8 Grease the seal face on the lower filter housing 5 wit...

Page 15: ...S cold water c 1 S cm at 600 S cold water and non exhausted ion exchanger Retention RO Nominal salt retention c 95 Membrane TFC bacteria resistant coil module Pre filter I Particle fine filter 10 5 m...

Page 16: ...tstra e 6 10 10829 Berlin Germany e mail info melag de Web www melag com Responsible for content Technical Office Subject to technical alterations Your stockist Rev 3 18 1685 Modification date 2018 08...

Reviews: