MELAG MELAdem 40 Operating Manual Download Page 15

 

Operating manual of the MELAdem

®

40

 

15 

3.8 

Connection of the MELAdem

®

40 to the Vacuklav

®

40-B / 44-B

___________________

 

The connection between the components of the system is provided by the hose delivered with the 
equipment (outside diameter = 6 mm; hose-wall thickness = 1 mm; MELAG article no. 28820). 

Important:  do not allow the plastic hose to be twisted or curled or mashed (compressed).

 

The MELA

dem

®

40 is connected to the autoclave according to the following diagram. 

Figure 13: 

Connection of MELAdem

®

40 to the Vacuklav

®

40-B / 44-B

 

 

Steps to follow for installation of MELA

dem

®

40 

1.  Make sure that the MELA

dem

®

40 has been correctly attached to the autoclave, and that the mixed-

bed-resin cartridges are properly installed. 

2.  Close the tap water line. 
3. 

Important Note: 

Start the Vacuum Test program. Stop this program after it has run approx. 20 

seconds. This will dispel the remaining water pressure in the system. 

4.  Unscrew the cooling-water feed hose (12) from the unit. 
5.  Screw the cooling-water branch fitting (11) into the unit. Screw the cooling-water feed hose (12) onto 

the cooling-water branch fitting (11). 

6.  Now attach the plastic hoses (3) to the MELA

dem

®

40 (2). Attach the feed hose (11) to the 

MELA

dem

®

40 to the left elbow fitting. Attach the output hose (3), from the MELA

dem

®

40 to the 

autoclave, to the right elbow fitting of the MELA

dem

®

40. 

7. Attach 

the 

MELA

dem

®

 feed-water filter (5) in-line to the plastic hose. 

8.  Make sure that all line connections are tight, and that there are no leaks. 
9.  Open the tap water line. 
 

If MELA

dem

®

40 is installed directly to the water tap, we recommend using a MELAG Waterstop valve (MELAG art-

no. 01056). In the event of leaks or overflow during operation, the moisture sensors of the Waterstop valve, 
installed on the floor, will activate a solenoid valve to shut off the tap water supply. 

 
With the spray pistol MELA

jet

®

 (1) instruments can be cleaned or rinsed after disinfection with 

demineralized water before sterilization. 

MELA

jet

 

(MELAG Art.-No.: 30300) optional 

MELA

dem

40 

Tap water supply hose to MELAdem

40 

Feedwater input fittings for demineralised water 
(MELAG article no. 53465 –delivered with 
MELA

dem

40) 

Feed water filter for MELA

dem

®

40 

Wall drain (nominal pipe size = NW 40, to be provided 
by the user) 

Dual-chamber waste-water trap (MELAG article no. 
26635) 

Double hose- support sleeve with anti-flooding flaps 
(MELAG article no. 37400: included with surface-
mounted wall drain [MELAG article no. 37410]) 

Cooling-water drain hose (DN 16 rating) 

10 

Cooling-water drain fitting 

11 

Cooling-water feeder and branch fittings (MELAG 
article no. 37241) 

12 

Cooling-water feed hose (pressure hose) 

13 

Water faucet (provided by the user) 

14 

Power connection 

15 

Optional:  leak monitor with shut-off valve and sensor 
(MELAG article no. 01056) 

16 

Rear view of Vacuklav

40-B / 44-B 

 

1

2

3

4

5

6

7

9

10 

11 

12 

13 

14 

15 

16 

17 

 

17 

Safety combination, consisting of a return-flow inhibitor 
(non-return valve) together with a backsiphonage pre-
venter, in accordance with EN 1717 

Summary of Contents for MELAdem 40

Page 1: ...s succeeded in becoming a leading manufacturer of sterilization equipment More than 420 000 MELAG sterilization units sold throughout the world testify to the exceptional quality of our products which...

Page 2: ...fe and water tight closing of the cartridge housings 21 6 TECHNICAL DATA 22 7 CONSUMABLES AND SPARE PARTS 22 Abbildungen Figure 1 View of the interior of the MELAdem 40 4 Figure 2 Incoming and outgoin...

Page 3: ...ected to a MELAG autoclave for example the MELAG Vacuklav Vacuquick or Euroklav models to automatically supply the autoclave with demineralized water The MELAdem 40 can however also be used as a compl...

Page 4: ...ift it vertically at the arrow markings 2 Filter housing wrench The filter wrench is used to open the cartridge enclosures It can be stored under the cover of the MELAdem 40 3 Fastening screws These 8...

Page 5: ...OF INSTALLATION CONNECTION PAGE Wall installation 3 3 Installation of the MELAdem 40 to a wall 6 Vacuklav 40 44 B 3 4 Installation of MELAdem 40 to the Vacuklav 40 B 44 B 7 Vacuquick 14 B 3 5 Install...

Page 6: ...m please use enclosed template for drilling Make sure that the wall can carry the weight of the system approx 2 4 kg Be sure to choose the proper fastening material in accordance with the wall for exa...

Page 7: ...ructions in Section 3 6 below These instructions tell you how to attach the mounting fixtures yourself These mounting fixtures are concealed under cover pieces that can easily be removed by hand see F...

Page 8: ...to the unit it is not necessary to remove the housing from the autoclave 2 Remove the cover pieces from the mounting fixtures as described above 3 Hang the MELAdem 40 down into the mounting fixtures...

Page 9: ...follow for installation of MELAdem 40 1 Warning First switch off the unit and pull out the power plug 2 Remove the housing cover of the autoclave 3 Press the mounting fixtures article no 37106 into th...

Page 10: ...pter plate 1 2 Figure 9 Correct position of the MELAdem 40 to the Vacuquick 13 B Vacuquick 14 B 1 Housing cover of the Vacuquick 13 B Vacuquick 14 B 2 Adapter plate Art No 79780 3 MELAdem 40 1 MELAdem...

Page 11: ...ing water network For protection of a drinking water network to which an autoclave is connected EN 1717 Part 4 stipulates that a combination of the following two units on the rear wall of the autoclav...

Page 12: ...n angle type non return valve and hose 10 mm already exists or will be especially installed Installation of an additional water faucet with integrated safety combination MELAG article no 58130 by dire...

Page 13: ...mentioned in the guide to installation in chapter 3 1 is made with plastic threaded connections The connection of the free hose ends to the water supply and the consuming device depending on the inst...

Page 14: ...le of the conductivity sensor to the control monitor connection Probe Adjusting the conductivity limit value on the control monitor 1 The housing lid is loosely positioned on the control monitor upon...

Page 15: ...plastic hose 8 Make sure that all line connections are tight and that there are no leaks 9 Open the tap water line If MELAdem 40 is installed directly to the water tap we recommend using a MELAG Water...

Page 16: ...return flow inhibitor non return valve together with a backsiphonage preventer in accordance with EN 1717 Figure 14 Connection of MELAdem 40 to the Vacuquick 14 B Steps to follow for installation of...

Page 17: ...o 01056 29 Untreated water input hose to MELAdem 40 10 19 20 5 6 7 9 12 11 21 23 24 25 26 22 27 28 31 29 31 Safety combination consisting of a return flow inhibitor non return valve together with a ba...

Page 18: ...ter for MELAdem 28 Input hose for distilled or demineralized water 29 Water faucet provided by the user 30 Optional leak monitor with shut off valve and sensor MELAG article no 01056 36 Safety combina...

Page 19: ...depending on water pressure from the tap This first filling is necessary to provide enough treated water for the autoclave or another consumer for example a MELAjet Before first putting into operatio...

Page 20: ...oclave then normal delivery of the water to the MELAjet will release the water pressure Remove the cover of the MELAdem 40 to gain access to the filter wrench Remove the wrench from the MELAdem 40 see...

Page 21: ...sing For fastening the cartridge housings you must only use the wrench 2 delivered with MELAdem 40 1 open off 2 3 Figure 18 How to use the cartridge housing wrench The safe and water tight closing of...

Page 22: ...ew fitting 90 1 4 with nut 61050 1 Cartridge housing wrench 16070 2 Hose clamp 37475 1 Clamping sleeve on order to fixing hoses at the MELAdem 40 bushing Connection sets and fittings for the various m...

Reviews: