User Manual for the Cliniklav
25
39
Warning message
Cause/Remedy
MELA
42 the red light should indicate 'power on'
•
The printer may be "Offline". Select "online"
(MELA
42, press "SEL" button, green LED "SEL"
should shine)
The internal protocol memory is full (40 cycles), a output
medium is registered, and in the menu Function
Batch
output
the option "Immed. output? NO" is selected. The
message is displayed when a program is started. Pressing
the "Start/Stop" key again deleted the message and the
program starts:
•
You can continue operations simply by pressing the
"Start/Stop" key twice when you start a program.
•
Select "Immed. output? YES" (see section 5.7.1)
•
Delete stored logs (see section 5.7.5), if necessary
print-out all stored cycle logs first (see section 5.7.4)
In the menu Function
"Batch output"
"Output
medium", select the "No output" option (see section 5.6)
The service message is activated after a certain number of
batches or a set operating period, when a service is due.
The message appears before the start of every program. If
you press the "Start/Stop" key again the message is
deleted and the program starts.
•
You can continue operations, by simply pressing the
"Start/Stop" key twice when you start a program.
•
Have a service carried out as recommended by an
authorised MELAG servicing company or your specialist
dealer.
The cycle counter for servicing should be reset during the
service.
The leak rate during the vacuum test exceeds the limit
value:
•
Check and if necessary clean the door seal and the rim
of the chamber
•
Repeat the vacuum test with a completely cold
autoclave
If no other malfunction messages occur during operation,
you can continue to use the autoclave until the regular
service, when the cause of the leak will be identified.
9.2 Malfunction messages
_________________________________________________
Malfunctions are generally reported by an "Error" on the display with the number of the error and its short
name.
Malfunction messages may occur without a program start (when the power is switched on or soon after), or
during a program.
If errors are reported during a program, then in addition to the error report the program will also be stopped.
This may be accompanied by the equilibration of the pressure in the autoclave, and in this case the error
message will alternate with the messages "Pressure release", or "Ventilation", and "End".
After the termination, the display will alternately show the error message and "Acknowledge with button ‘-‘ "
and then "Stop/End ". Pressing "-" deleted the error message (if the error is not permanent). Until you have
quit the malfunction message the autoclave door cannot be opened. If a program has been prematurely
terminated in this way the autoclave load must always be regarded as being
not sterilized
. We recommend
that you unload the autoclave, carry out a sterilization cycle without any load (the drying may be impaired for
this first cycle) and then reload the autoclave and repeat the interrupted operation cycle.
If an output medium is connected and "Immed. output? YES" is selected, a log will automatically be printed
out at the end of the termination.
The log output shows the full name of the error, and if a program has been interrupted before completion it
will also show "Load not sterile". The following list gives malfunction messages, the cause and possible
remedies.
P r ot o co l m emo r y
f u l l
E x ce c ut e s er vi ce
p l ea se
T e s t u ns u cc es s f ul
L e a k r a te : 3. 2