11 / 18 

9756269 

 

Safety instructions 

 

 

 

 
TopCut  is  necessarily  equipped  with  a 

sharp  blade

  at  the  front.  When  the 

cutting mechanism is in the idle position, the blade is covered and only revealed 
when  the  door  to  the  wash  chamber  is  closed. 

If  the  TopCut  is  directly 

operated by hand, the blade poses a risk of injury. 

 

 

 

If the TopCut is bent from improper use, the blade may remain uncovered after 
use. In this case, the TopCut is to be disabled immediately and secured against 
use. 
 

 

 

Insert the bags into the holder, do not throw them in as this could cause the door 
to close unexpectedly. 
 

 

 

If the bag is not correctly cut open, place it centrally in the holder and cut it open 
again. If the TopCut still fails to cut it, you must inform an authorised member of 
staff. 
 

 
 

Once the chamber door has been closed, do not reopen it. Start the programme 
immediately. 
 

 

 

During servicing:

 use caution when replacing the blade: 

risk of injury.

 

 

 

Please use the cleaning and disinfection machine as instructed. If it is used or 
operated  differently,  damage  and  risks  may  occur  for  which  we  accept  no 
liability. 

 

 

 

 

 

Wear  personal  protective  equipment  (rubber  gloves).  In  order  to  avoid 
contaminating  the door handle and the  work area, gloves should be removed 
after  loading  the  TopCut.  Disinfect  your  hands  afterwards.  Observe  the 
operator's hygiene regime. 

 

 

 

Please read and observe the safety instructions and the operating instructions 
for the cleaning and disinfection machine. 

 

 

 

  

 

Before  removing  the  TopCut,  check  whether  the  blade  has  been  properly 
retracted  and  ensure  that  the  holder  is  unsoiled.  Once  the  TopCut  has  been 
dismounted,  carry  it  on  the  side  facing  the  cutting  mechanism.  Do  not  use 
moving parts for support. Handle the TopCut with care. Do not throw, swing or 
drop it. 
 
TopCut  is  not  to  be  used  in  devices  with  an  automatic  door.  If  used  with  an 
automatic door, the increased strain on the door drive may be damaged, as may 
the TopCut itself.  

 

 

 
For your safety and to avoid danger, 

only

 use authentic MEIKO spare parts.

 

 

Summary of Contents for TopCut H

Page 1: ...pdate 2021 03 www meiko global com Operating instructions Universal vessel holder with urine and secretion bag cutter TopCut H Translation of the Original operating instructions Read operating instruc...

Page 2: ...TENDED USE 3 2 1 Suitable bag types 4 3 STRUCTURE 7 4 FUNCTIONAL DESCRIPTION 8 5 OPERATION INSERTING THE BAG 9 6 CLEANING 10 7 SAFETY INSTRUCTIONS 11 8 NOTES IF TOPCUT IS NOT WORKING PROPERLY 12 9 MAI...

Page 3: ...the scope of supply concerning the design equipment and technical features at any time No claims may therefore be based on the details the images or the descriptions contained in these operating instr...

Page 4: ...0 Dianeal PD1 1500 Bbraun Ureofix 1500 3500 Ureofix with outlet 2000 Secretion bag 1500 2000 Vacufix 1000 Curion CuriBag 1500 Dahlhausen Urine bag with tube 1500 Gastric suction bag 1500 Kendall Curit...

Page 5: ...5 18 9756269 BBraun Ureofix BBraun secretion bag Worek with outlet BBraun Vacufix...

Page 6: ...the TopCut H Commode pot with lid and a urine bottle together Plastic commode pot with lid and a urine bottle together Bedpan with lid and a urine bottle together 3 urine bottles together Urine colle...

Page 7: ...7 18 9756269 3 Structure Locking lever TopCut H Blade Bag feed Vessel holder...

Page 8: ...used as a replaceable holder The cutting mechanism is activated as the wash chamber door is closed The bag is cut open in the middle 30 to 40 mm and then the bag empties itself with the wash chamber...

Page 9: ...of injury Inserting the bag to be emptied Open the locking arm towards the back place the bag centrally in the bag tray in front of the blade Close the locking arm If there is a tube place it under t...

Page 10: ...programme has finished the bag has been emptied and disinfected and can be removed 6 Cleaning Once the programme has finished the bag has been emptied and disinfected on the outside The TopCut H has...

Page 11: ...e blade risk of injury Please use the cleaning and disinfection machine as instructed If it is used or operated differently damage and risks may occur for which we accept no liability Wear personal pr...

Page 12: ...its operational state or if appropriate replace it The retaining plate is not screwed onto the bracket on the TopCut The TopCut is pushed upwards when the door is closed An authorised member of staff...

Page 13: ...oving housing parts and when working inside the machine Before removing the TopCut check whether the blade has been properly retracted and ensure that the holder is unsoiled Once the TopCut has been d...

Page 14: ...et on the wash chamber door and screw in 2 Check working load filled urine bag perhaps fill with water into TopCut holder and close door to wash chamber 11 Decommissioning If the machine has been oper...

Page 15: ...15 18 9756269 12 Spare parts list Bag cutter TOPCUT H 12 8 9 10 1 6 4 3 12 5 11 12 10 15 15 14 2 13 7...

Page 16: ...LE ROLLER POULIE 6 9658626 MESSER KNIFE COUTEAU 7 9658624 WELLE 8X36 SHAFT 8X36 ARBRE 8X36 8 9658893 SCHENKELFEDER LEG SPRING RESSORT BRANCHES 9 9658900 SCHENKELFEDER LEG SPRING RESSORT BRANCHES 10 96...

Page 17: ...17 18 9756269...

Page 18: ...Valid from 2009 09 Update 2021 03 MEIKO Maschinenbau GmbH Co KG Englerstra e 3 77652 Offenburg Germany www meiko global com info meiko global com Design and construction subject to change without prio...

Reviews: