background image

OPERATING INSTRUCTIONS

OPTICAL SET-UP AND BRIGHTFIELD ILLUMINATION

After connecting the illuminator plug to the receptacle

at the back of the microscope base, please

follow as indicated below: 

[1]  Turn on the illuminator.

Place the specimen slide you wish to examine on the microscope stage

and rotate the nosepiece to bring the 10X objective into position for focus.

[2]  Make sure that both the field iris
(a) and the aperture iris (b) are fully
open.

[3] Slide the Illumination Selector to the central position where the middle Empty Hole comes 
across the Light Axis.  This is the Brightfield Illumination.

ILLUMINATION SELECTOR 

[4] Pull the Analyzer Slide 

out to the full so the Analyzer gets out of the light path.

[5]  Focus down on your specimen slide until surface detail can be seen.  Adjust the brightness of the
built-in light source, using the intensity control knob, left-hand back on the base. 

[6]   BINOCULAR ADJUSTMENT
Comment:  Using a binocular body is much more efficient and less tiring than monocular bodies, but it
must be adjusted correctly.  When it is perfectly adjusted the images coming from the two eyepieces are
“fused” into one better image in eyes of the observer.

6

3

Receptacle 

SELECTOR LEVER

Polarizer

Empty hole for Brightfield

Darkfield stopper

SLIDE-IN ANALYZER 

The Analyzer

is mounted in an in-tube slider which moves the analyzer in and out of the optical path

and rotates between 0

and 90

by the lever

.  When “in” and at  45

position, and with the

Polarizing filter in position, these elements are said to be “crossed” and the field of view is said to be

“extinguished”.  In this condition the
field of view is dark - except for
optically active elements in the field,
which rotate the angle of polarization
and thus become visible against a
dark background. 

TRANSMITTED LIGHT ILLUMINATION

To cover observation of both transparent and semi-transparent specimens, this series of metallurgical

microscopes are provided with a set of transparent glass stage plate, substage condenser and a semi-

Koehler type 6V 30W Halogen lamp illuminator.  The usage of the Transmitted Light illuminator is as

follows: 

[1] Place the specimen at the center of the stage plate, and switch on and control the light by simply

rotating the Intensity Control knob

[2] Lift up the substage condenser to the utmost by the Condenser Focusing Control

and then open

up the irises for Field and Aperture.

[3] Focus down on the specimen, and also focus on the closed Field Iris itself. These two focused

images of the specimen and the closed Iris are seen overlapped.  Then, open the Field Iris

slowly

until the field of view comes to cover the whole image of the specimen.  Reduce the Aperture slowly

until you can view the image in the best appearance.  Too much reduction of Aperture Iris is not good

as the resolution capacity gets much less due to the roughened image.  (By the way, the closed Field

Iris can not be seen through 100X objective.)

Analyzer slide 

Analyzer rotating

lever

Tension control knob 

Condenser Focusing
Control 

Field Iris & Control 

Condenser centering  

screw

Aperture Iris & Control 

lever

Intensity Control knob 

Analyzer 

Slider

Summary of Contents for ML 8500 Series

Page 1: ...JAPAN...

Page 2: ...CONTROLS ON OFF AND INTENSITY CONTROL FOR TRANSMITTED ILLUMINATOR ILLUMINATOR CONDENSER MICROSCOPE BASE WITH BUILT IN TRANSFORMER FIELD IRIS CONTROL FOR TRANSMITTED LIGHT a Field Iris Control Lever B...

Page 3: ...the clamp screw Now the binocular or trinocular body can be mounted on the vertical illuminator To mount the body loosen the clamp screw and insert the cone fitting of the body into the recess on the...

Page 4: ...comes across the Light Axis This is the Brightfield Illumination ILLUMINATION SELECTOR 4 Pull the Analyzer Slide out to the full so the Analyzer gets out of the light path 5 Focus down on your specime...

Page 5: ...n the dimension set may not in this case exactly correspond to the window s indication 10 Now turn the field iris adjustment lever a until the field iris is seen in the field of view 11 Your light sou...

Page 6: ...stopper comes across the Light Axis centrally 2 Open up both irises for Field a and Aperture b 3 Focus down on the specimen just in the same way as for the Brightfield observation The image is observe...

Page 7: ...ight illuminator is as follows 1 Place the specimen at the center of the stage plate and switch on and control the light by simply rotating the Intensity Control knob 2 Lift up the substage condenser...

Page 8: ...of the light to the binocular eyepieces In the system the MA150 50 or MA150 60 Camera Attachment should be used with the SLR camera of your choice Please note that one of the large range of T2 Adapto...

Page 9: ...rimented with in order to obtain the best monitor image MAINTENANCE AND CARE BULB REPLACEMENT ON REFLECTED LIGHT ILLUMINATOR When changing light bulbs in the illuminators always disconnect the plug fr...

Page 10: ...ll times in order to prevent dust from falling on the internal optics Store the microscope in a safe clean place when not in use for an extended period of time CLEANING Clean exposed lens surfaces car...

Page 11: ......

Reviews: