user manual V3.1
- 41 -
Element S-parameters
Several Cal Kits are provided with the S-parameter files of the Elements. These should be
the
raw S-parameters
without any deembedding for length. These S-parameters must of
course be measured against a better calibration kit.
The S-parameters can be imported from a Touchstone format file (.S1P or .S2P), version
1.0.
After browsing a Touchstone file the screen shows the properties of the file. With a two-
port file you can select the desired data set (S11, S22).
It is recommended to have 10MHz or less between sample points. The optional Resample
feature can be used to limit the frequency range or decrease the number of samples to
store.