7. 

BATTERY REPLACEMENT

 

 

In the event the battery requires replacement, it can be replaced with an Endevco model EBT26.  It can be 
installed by Endevco or the user can replace the battery.  The battery is located in the upper right corner in 
figure 3. 
 
Below are step-by-step instructions to follow for replacing the battery: 
 

1.

 

Remove the four screws on the bottom of the instrument.  These screws may be located under the 
rubber feet. 

2.

 

Gently remove the top cover and the battery will be visible (see figure 3).  The battery will lift out 
since it is held in-place when compressed by the upper and lower case. 

3.

 

It is recommended that the wires be cut close to the battery leaving the other end of the wires 
attached to the circuit board and plug.  Strip the ends of the wire 

4.

 

Strip the ends of the leads on the new battery.  

5.

 

Solder the new battery leads to the leads that were left in the simulator.  Cover the solder joint with 
shrink tubing 

6.

 

Reassemble the case and check for proper operation. 

 

CALIBRATION AND TESTING 

 

Annual calibration is recommended to ensure accuracy and proper performance.  Calibration service is 
available from the factory, or the user can perform the calibration by following the following procedures. 
 
 

8.1 

Test Equipment 

 

Unless otherwise stated, the following equipment is required.  Items may be substituted with equipment of 
similar performance

 

 

DC Power Supply, 9 Volts regulated.  The power supply provided with the 4830A is preferred 

 

Oscilloscope 

 

Frequency counter 

 

RMS Voltmeter 

 

Isotron Signal Conditioner model 4416B 

 

Single/differential charge amplifier model 6634C 
 

8.2 

Testing 

 
8.2.1

 

With the unit switched on, verify that the Battery Low message is not displayed.  Use the 9 V 
regulated power supply to recharge the batteries, if required.  The procedure may be performed 
with the power supply connected. 
 

8.2.2 

Turn the unit off and connect the 9 Volt power supply (note that the center pin is +).  The unit should 
automatically turn on with the power supply connected and charging the internal batteries.  Verify 
that the yellow LED charge indicator is lit when the power supply is connected. 
 

Summary of Contents for ENDEVCO 4830A

Page 1: ...4830A Accelerometer simulator Instruction manual IM4830A Revision E1...

Page 2: ...DOCUMENT SUBJECT TO U S EXPORT CONTROLS COMPLIANCE APPROVIAL REQUIRED PRIOR TO DISTRIBUTION IM4830 Page 2...

Page 3: ...ble Simulator Acceleration Displacement Velocity Tachometer Output signals are provided that simulate voltage mode accelerometers IEPE ISOTRON etc and charge mode accelerometers both single ended and...

Page 4: ...ED will extinguish The orange trickle charge LED will remain lit indicating that the battery voltage is being maintained topped up by a trickle charge current This configuration is recommended for opt...

Page 5: ...o increase value or down to decrease value note that the values either side will change when roll over occurs depending on count up or down The cursor has to be returned to right hand area to select a...

Page 6: ...gnal conditioners The connectors provide the following outputs TACHO Provides a simulated tachometer signal with TTL 5 Volt pulses ICP Provides a millivolt output signal as selected on the panel This...

Page 7: ...QUIRED PRIOR TO DISTRIBUTION IM4830 Page 7 The rear panel contains the power switch and charger input The LED s labeled F and C indicate fast charge F and trickle charge C The charge mode is selected...

Page 8: ...DOCUMENT SUBJECT TO U S EXPORT CONTROLS COMPLIANCE APPROVIAL REQUIRED PRIOR TO DISTRIBUTION IM4830 Page 8...

Page 9: ...DOCUMENT SUBJECT TO U S EXPORT CONTROLS COMPLIANCE APPROVIAL REQUIRED PRIOR TO DISTRIBUTION IM4830 Page 9...

Page 10: ...DOCUMENT SUBJECT TO U S EXPORT CONTROLS COMPLIANCE APPROVIAL REQUIRED PRIOR TO DISTRIBUTION IM4830 Page 10...

Page 11: ...the solder joint with shrink tubing 6 Reassemble the case and check for proper operation 8 CALIBRATION AND TESTING Annual calibration is recommended to ensure accuracy and proper performance Calibrati...

Page 12: ...io settings 1 0 10 0 and 100 0 Limits are 1 Verify Tach Output is 5 Volts peak to peak 8 4 Amplitude Accuracy mV OUTPUT 8 4 1 Connect a true RMS voltmeter and an oscilloscope to the mV output socket S...

Page 13: ...RT CONTROLS COMPLIANCE APPROVIAL REQUIRED PRIOR TO DISTRIBUTION IM4830 Page 13 444 0 313 96 310 8 317 1 555 0 392 44 388 5 396 4 666 0 470 93 466 2 475 6 777 0 549 42 543 9 554 9 888 0 627 91 621 6 63...

Page 14: ...TRON ICP Output With the exception of the Tach all other outputs of the 4830A are derived from the mV output To verify their operation suitable calibrated signal conditioners should be used e g for th...

Page 15: ...8 to 1 02 8 7 Amplitude Accuracy DIFF Charge Output 8 7 1 Connect the DIFF Charge Output to the input of the Model 6634B Signal Conditioner Set the 6634B for Type DIFF Sens 10 00 FS 20 0 Mode g DC pK...

Page 16: ...ransient power glitches The problem can generally be corrected by doing a Memory Reset as described in item 4 If the problem persists the unit is most likely defective 10 0 TECNICAL SUPPORT For techni...

Reviews: