background image

13

RS232 / USB connection

Data can be transferred to a PC via an RS232 or USB port. If using the 
RS232 port, use the null modem cable supplied. The RS232 settings 
are 38400 Baud, 8 data bits, 0 parity, 1 stop bit, no flow control 
(handshake).
If using the USB port, ensure that the USB driver supplied on the 
accompanying product CD has been installed 

BEFORE

 connecting the 

instrument. Installation instructions are also to be found on the CD.

Programmes such as Megger Download Manager may be used to 
download the results stored in memory. Programmes such as Microsoft

® 

HyperTerminal may be used to record real time data.

Battery bar graph

This is a battery symbol on the LCD display comprising 4 pairs of 
segments. The battery is monitored continuously when the instrument 
is turned on. The charge remaining in the battery is shown in the table 
below.

Fully charged battery 

 

 

50% charged battery  

Tests cannot be started, and the battery  
may fail at any time.  

 

Symbol flashes when there is not 
enough charge for a test. The 
instrument then turns itself off.  

When mains power is present the indicator shows the battery is being 
charged by animating the segments of the bar graph.

Voltage at terminals

The test voltage will default to that selected in the previous test. If the 
instrument has been switched off since the previous test, the test voltage 
defaults to 250 V.

If there is an external voltage greater than 50 V present, this will be 
displayed regardless of changes made to the test voltage. In this case the 
instrument will not perform a test. The high voltage warning indicators 
flash, and the beeper sounds to warn of the hazard until the external 
voltage becomes less than 50 V.

When testing, the voltage displayed is the voltage present at the test 
terminals of the instrument. If the test voltage is changed, by pressing 
either of the test voltage 

s

 and 

t

 buttons during a test, the new test 

voltage is displayed momentarily.

After testing, the voltage displayed is the voltage present on the 
terminals. To see what the voltage was immediately prior to the end on 
the test, press either test voltage 

s

 or 

t

 buttons.

Timer indicator

The timer shows minutes and seconds. At the start of a test the timer 
will start from zero and at the end of a test it will stop. The duration of 
the last test remains on the display until another test is started.

Digital display

The digital display shows the resistance or current being measured 
during a test. The 

Ω

/I button toggles between the two. After a test the 

Summary of Contents for S1-552

Page 1: ...M S1 552 2 5kV Digital Insulation Tester USER MANUAL...

Page 2: ...turned off and the circuit discharged manually n Test leads including crocodile clips must be in good order clean and with no broken or cracked insulation n The instrument should not be used if any p...

Page 3: ...circumstances Failure to follow these instructions could result in damage to the instrument and or a hazard to the operator The word should is used to indicate that the instructions indicate best prac...

Page 4: ...retrieval facilities Polarization index and dielectric discharge tests are performed automatically and test duration and voltages can be adjusted according to user preference for these tests A step v...

Page 5: ...e so the ground connection will not be used Power lead connection table Connection UK International USA Earth Ground Yellow Green Green Neutral Blue White Live Line Brown Black If using a fused plug e...

Page 6: ...dicator Alarm Data recording Test modes Voltage at terminals Battery level Breakdown indicator High voltage warning indicator High voltage warning indicator Analogue display Timers Main T1 T2 Dielectr...

Page 7: ...isplay will depend upon the test mode selected whether timers T1 and T2 have been set and Power On Off button The instrument will only turn on if this button is pressed held and then released when the...

Page 8: ...nd current In the IR PI and DD modes the secondary display initially shows the PI polarity index DAR dielectric absorption ratio and on completion of the test the TC time constant and capacitance meas...

Page 9: ...ns timer defaults to 10 minutes as this is required for a PI reading Figure 3 The result of a dielectric discharge test DD Test settings T1 and T2 times set in order to measure the DAR Test condition...

Page 10: ...ct the test mode Modes of test to be chosen from include an insulation resistance IR test a polarization index PI test a step voltage SV test and a dielectric discharge DD test The insulation resistan...

Page 11: ...t now displays clr to indicate clear mode Press and hold down test button until clear process begins shown by analogue display lighting anti clockwise Press the MODE button to exit without deleting th...

Page 12: ...measured in the terminal Guard terminal For basic insulation tests and where there is little possibility of surface leakage affecting the measurement it is unnecessary to use the guard terminal i e if...

Page 13: ...s of the bar graph Voltage at terminals The test voltage will default to that selected in the previous test If the instrument has been switched off since the previous test the test voltage defaults to...

Page 14: ...Increase decrease the decrease the decrease the decrease the test voltage in test voltage in test voltage in test voltage in major steps major steps minor steps minor steps IR only IR only Vs AND Vt...

Page 15: ...provide accurate measurements with up to 4 mA of noise current present A level of noise exceeding the instruments specified capability will be indicated by an alternating sine wave on the display The...

Page 16: ...ent symbol PI On completion of the test the instrument will display the insulation resistance recorded at 1 minute and 10 minutes under the segment symbols 1m and 10m 16 TEST MODE SUMMARY Pressing the...

Page 17: ...ance threshold between the limits of 10 k and 15 T Release the function button to save the current limit Adjusting the alarm threshold level automatically activates the alarm limit mode This is indica...

Page 18: ...that the test leads are reasonably clean and dry The guard lead can be used to remove the effects of surface leakage if necessary When measuring resistances above 100 G the test leads should not be al...

Page 19: ...ths may occur Alternatively screened leads are available as an optional accessory from Megger The lead to the negative terminal is fully screened The screen is plugged into the Guard terminal divertin...

Page 20: ...n test voltage 20 SPECIFICATIONS Capacitance measurement accuracy 23 C 5 5 nF Voltage output accuracy 0 C to 30 C 4 0 10 V of nominal test voltage at 1 G load Current measurement range 0 01 nA to 5 mA...

Page 21: ...User guide on CD ROM 2000 213 RS232 cable 25955 025 UBS cable 25970 041 Optional Accessories HV test lead sets 3 x 3 m with un insulated small clips 8101 181 3 x 8 m with un insulated small clips 810...

Page 22: ...Tel 44 0 1304 502 243 Tel 1 610 676 8579 Fax 44 0 1304 207 342 Fax 1 610 676 8625 Megger operate fully traceable calibration and repair facilities ensuring your instrument continues to provide the hi...

Page 23: ...ry charge level indicator will need to calibrated to the new battery s characteristics End of life disposal WEEE The crossed out wheeled bin placed on the Megger products is a reminder not to dispose...

Page 24: ...3R 6A8 Canada T 1 416 298 9688 Canada only T 1 416 298 6770 F 1 416 298 0848 E casales megger com Megger products are distributed in 146 countries worldwide This instrument is manufactured in the Unit...

Reviews: