DAR is the ratio between the insulation resistance values at

30seconds (assigned t1) and at 60 second interval (assigned t2). i.e.

after 30seconds and 60seconds.

DAR = 60 second value/30 second value

During all insulation tests the symbol 

G

will flash indicating that a

test voltage is present.

(a) Insulation resistance testing-timed procedure.

1. Connect the test probes to the isolated circuit under test.

2. Turn the instrument ‘ON‘ by rotating the selector switch to the

required test voltage position.

3. Select the timed test (t) by pressing ‘PI/DAR/t’ function button

repetitively until the desired function is displayed.  

4. Once selected, press and hold the [TEST] button to start the test.

5. To abort the test early, press TEST or LOCK [

] buttons. 

6. At the end of the tests, the voltage will be discharged.

(b) Insulation resistance testing – PI and DAR 

The PI test will run for a period of 10minutes. After one minute a
test result is stored (t1). After 10 minutes a second test result is
stored (t2). The resultant ratio is then displayed on the screen.
The same procedure applies for the DAR timed tests, however the
test duration is 60 seconds, with the first result (t1) taken at 30
seconds and the second at 60 seconds (t2).

Results can be recalled to screen using the uA/v/s [

t

] keys. 

Note: DO NOT press PI/DAR/t [

s

] as it will change the test mode

and erase the current results.

Continuity testing [

] and buzzer [

]

1. Turn the instrument ‘ON’ by rotating the selector switch 

to 

the desired 

position.

2. If required the test lead resistance can be set to Zero (null) by

shorting the test leads together and pressing TEST. The null 

[

z

] symbol will show when this has been achieved and the

display will read 0.00 

.

3. Press the [

] button to enable/disable the audible buzzer

function. When enabled, the sounder symbol will be shown on

the screen display. The pass threshold is set to 2 

by default,

but is adjustable, as defined in Setup. 

Note that the buzzer defaults to OFF on power-up.

4. Connect the test leads to the isolated conductor(s) under test.

Summary of Contents for MIT415

Page 1: ...NSW 2125 Australia T 61 0 2 9659 2005 F 61 0 2 9659 2201 E ausales megger com Megger Limited 110 Milner Avenue Unit 1 Scarborough Ontario M1S 3R2 Canada T 1 416 298 9688 Canada only T 1 416 298 6770...

Page 2: ...M MIT415 Insulation and continuity tester User Guide...

Page 3: ...f it is damaged n All test leads probes and crocodile clips must be in good order clean and with no broken or cracked insulation n Ensure that hands remain behind guards of probes clips when testing n...

Page 4: ...st lead probes also observe voltage discharges to a safe level On battery cover see section 2 0 notes 10 and 11 At terminals do not exceed rated input voltage t Equipment protected throughout by Doubl...

Page 5: ...re both AC and DC components of the supply voltage AC DC In DC mode only the DC component is measured 1 Rotate selector switch to the V position 2 Connect test leads to the circuit under test 3 Press...

Page 6: ...displayed on the secondary display 5 The insulation test can be locked on by pressing the lock button whilst holding down the TEST button To disable lock press the TEST button or lock button 6 By pres...

Page 7: ...a test result is stored t1 After 10 minutes a second test result is stored t2 The resultant ratio is then displayed on the screen The same procedure applies for the DAR timed tests however the test du...

Page 8: ...ent firmware revision is displayed followed by the buzzer alarm threshold Displayed Meaning Default Setting symbol setting options BUZ Set top threshold for continuity buzzer in ohms 2 1 2 5 10 20 Buz...

Page 9: ...r and lift out the cells Ensure that the replacement cells are fitted with the correct polarity in accordance with the label in the battery compartment Replace and re secure the battery compartment co...

Page 10: ...t on load 1 mA at min pass value of insulation specified in BS7671 HD384 and IEC364 EN61152 2 2 mA max EN61557 operating range 0 10 M to 1 00 G Leakage current range 10 A to 2000 A Leakage current 10...

Page 11: ...r Operates at 25 volts a c or d c on any range except OFF Frequency 15 400Hz 15Hz 99 9Hz 0 5 1 digit 100Hz to 400Hz Power supply 5 x 1 5V cells type IEC LR6 AA MN1500 HP7 AM3 R6HP Alkaline NiMH rechar...

Page 12: ...ture range 30 C to 80 C Calibration Temperature 20 C Maximum altitude 2000 m Dust and water protection IP54 Accessories Includes accessories 2 wire test lead set and crocodile clips 6220 813 SP5 Remot...

Page 13: ...t is necessary to retun an instrument for repair a Returns Authorisation number must first be obtained by contacting one of the addresses shown You will be asked to provide key information such as the...