4. To start the test press the TEST button.
Resistance and current values continually up date on screen.
The current starts at 1 mA and can reach a maximum of 1 A if a
higher current is required for the resistance being measured.
End of the test
To stop the test press the TEST button.
Note:
The TICK button (
) will set a new reference value.
Warning: Do not disconnect test leads until discharge is completed.
During the discharge phase, the buttons and rotary switch will be inactive. While discharging is still
in progress a warning symbol (
F
) will flash on the left of the screen and an audible warning buzzer
will sound. Do not remove the test leads until discharge is complete and the warning disappears.
5.1.2 Second test
1. To start the test press the TEST button.
The animated chevron sequence will be displayed during the
pre-charge phase of the measurement, to be replaced by
resistance and current values updating as the instrument
auto-ranges.
Readings continue to update until a stable measurement is
achieved.
2. When the result is stable press the TEST button.
The Difference Meter records the reading.
End of the test
To stop the test press the TEST button.
Note:
The TICK button (
) will set a new reference value.
Warning: Do not disconnect test leads until discharge is completed.
During the discharge phase, the buttons and rotary switch will be inactive. While discharging is still
in progress a warning symbol (
F
) will flash on the left of the screen and an audible warning buzzer
will sound. Do not remove the test leads until discharge is complete and the warning disappears.
Note:
If continuity is lost from either the C or P connection
during a test, this screen will be displayed for 3 seconds.
The DLRO2 will then return to the start of the test.
Tests, Inductive mode