Megger DCM340 User Manual Download Page 3

Specifications 

1-1  General Specifications 

 

LCD display digits :  

3 3/4 digit large scale LCD readout. 

Display count : 

4000 counts. 

Measuring rate : 

1.5 times / sec. 

Overrange display :  

“OL” is displayed for “

Ω

”  functions,  

shows the real value for “A” and “V” 
function. 

Automatic power off time :  

Approximately 30 minutes after 
power on.  

Low battery indicator :  

 is displayed. Replace the battery 

when the indicator 

 appears in the  

display.

 

Power requirement :  

9V PP3 / 6LR61 battery. 
Rechargeable batteries are not  
suitable for use with this instrument. 
 

1-2  Environmental Conditions 

 

Indoor Use. 
Calibration :  

One year calibration cycle. 

Operating temperature :  

0°C ~ 30°C (

80% RH)  

30°C ~ 40°C (

75% RH)  

40°C ~ 50°C (

45% RH) 

Storage temperature :  

-20 to +60°C, 0 to 80% RH (batteries 
not fitted).

 

Overvoltage category :  

IEC 61010-1 600V CAT.

CAT.

 equipment is designed to   

protect against the transients in the  
equipment in fixed installations, 
such as distribution panels, feeders  
and short branch circuits and light- 
ing systems in large buildings. 

Operating altitude : 

2000m (6562 ft)

 

Conductor Size : 

35mm diameter.

 

Pollution degree : 

2  

EMC : 

EN 61326-1

   

 

Function 

Range 

Accuracy 

V

 

0~400.0Vrms 

400~600Vrms 

 

±

(1.0%+ 5 dgt) 

 50Hz ~ 500Hz 

V

 

0~400.0 V 

400~600 V 

 

±

(0.7% + 2 dgt) 

Overload protection : 

600V rms 

Input impedance : 

1M

Ω

 // less 

than 100pF. 

AC Conversion Type : 

AC Conversion are average sensing 
rms indication calibrated to the rms 
value of a sine wave input.

 

Resistance & Continuity

 

Function  

Range 

Accuracy 

Ω

 

 

400.0

Ω

 

±

(1% + 3 dgt) 

Overload protection : 

600V rms

 

Max. open circuit voltage : 

3V 

Continuity check : 

Internal sounds  

activates if the resistance of the  

circuit under test is less than 30

Ω 

approximately.

 

AC Current

 

*1 : 

Over 80% of full scale add 

±

1.6%

 

Overload protection : 

600A rms

 

AC Conversion Type and additional 
accuracy is same as AC Voltage.  

Position Error : 

±

1% of reading.

 

Frequency : Hz

 

Range  

Resolution 

Accuracy 

20~400Hz 

1Hz 

±

(0.1% + 2 dgt) 

Min/Max Hold : 

add 

±

 15 dgt to  

accuracy for ACA / DCA. 
* Automatically switch to the low  
  resolution range at Peak Hold  
  and MIN MAX Hold.  

Function  

Range 

Accuracy 

A

(50~60Hz) 

 

0~400.0A 

±

(1.5%+ 5 dgt) *1  

  400~600A 

±

(2.0%+ 7 dgt)  

A

(61~400Hz) 

 

0~400.0A 

±

(2.0%+ 5 dgt) *1 

  400~600A 

±

(2.5%+ 7 dgt)  

DC Current

 

Function  

Range 

Accuracy 

   A

 

0~400.0A 

 

±

(1.2%+ 5 dgt)  

400~600A 

 

±

(1.9%+ 7 dgt)  

Overload protection : 

600A rms

 

Position Error : 

±

1% of reading. 

Addition error according to rema-
nence 

: 1% max. of current crest. 

 

Auto Power Off (APO) 

The meter will automatically shut itself 
off after approximately 30 minutes  
after power on.

 

Maintenance 

Do not attempt to repair this Meter. 
It contains no user-serviceable 
parts. Repair or serving should only  
be performed by qualified personal.

 

Cleaning 

Periodically wipe the case with a  
dry cloth and detergent do not use  
abrasives or solvents.

 

Shock vibration :  

Sinusoidal vibration per MIL-T- 
28800E (5 ~ 55 Hz, 3g maximum). 

Drop Protection : 

1.2m drop to 

hardwood on concrete floor. 

 

1-3  Electrical Specifications 

Accuracy is 

±

(% r number  

of digits) at 23°C  

±

 5°C < 80%RH. 

Temperature coefficient :  

0.2 x (Specified accuracy) / °C,  
< 18°C, > 28°C . 

 

Voltage

 

Overload protection : 

 600A rms.

 

Sensitivity : 

3A rms for ACA (A

)

(>400Hz Unspecified) 

Peak Hold : 

±

 (3% + 15dgt) 

*>600Vpeak Unspecified. 