background image

Live Circuit Warning
When more than 25V is applied to the terminals in the insulation

ranges, the instrument defaults to a voltmeter will give an

audible warning if a test is attempted. On all other switch

positions except OFF/v/mV when more than 10V is applied the

default voltmeter will be activated. Testing will be inhibited. 
Voltage Testing on High Energy Systems
Use extreme care when using or measuring voltages above

30V, particularly in high energy systems. Fused test leads are

available as optional accessories for local situations where

increased protection is required.
Auto-shut Off
To conserve battery life, Auto-shut Off (preceded by a series of

bleeps) operates after approx. 10 minutes of instrument

inactivity on the insulation ranges, 5 minutes on all other ranges.

If the instrument is switched on whilst holding the

key,

the Auto shut-off time is extended to 60 minutes. To restore

operation after Auto-shut Off, either select OFF followed by the

required switch position.
Note: It is recommended that the instrument is switched to the
OFF position when not in use. 

Backlight
The backlight is activated by pressing the

key. The

backlight will remain illuminated for approximately one minute

before automatically switching off to conserve battery life,

alternatively the

key can be re-pressed.

Insulation Tests (MΩ) (See fig.1)
The insulation tests apply a known voltage to the circuit under

test and measure the resulting leakage current. The circuit

under test must be completely de-energised and isolated

before test connections are made.
Insulation tests are only initiated when the TEST button is pressed.
1.  Set the range switch to the test voltage required.
2.  Connect the test leads, first to the instrument, and then to

the isolated item under test.

3. Press the TEST button to activate the test voltage. Take

the reading.

4. Release the TEST button at the end of the test. The last

reading will hold on the display.

5.  Any capacitive circuits charged during a test will automatically

discharge. If significant voltage remains the voltage warning

will occur and the voltage present displayed.

6.  Remove the test leads only when no voltage is indicated.
Locking Test Button (ltb)
When it is desired to do a long insulation test, the test can be

Summary of Contents for BMM80 Series

Page 1: ...ve no broken or cracked insulation Ensure that hands remain behind guards of probes clips when testing U K Safety Authorities recommend the use of fused test leads when measuring voltage on high energ...

Page 2: ...he instrument enables the optional Megger SP6F Switched Probe to be used for two handed probe operation The 250V 500V and 1000V ranges can be used to test electrical installations in compliance with B...

Page 3: ...strument is switched to the OFF position when not in use Backlight The backlight is activated by pressing the key The backlight will remain illuminated for approximately one minute before automaticall...

Page 4: ...Testing and value assessment can be found in Megger Limited publications listed in the Accessories page Automatic Discharge When the TEST button is released after an insulation test or re pressed if...

Page 5: ...ossible sources of error Measurements and results can be effected by the following The impedance of operating circuits connected in parallel Impedance such as inductors that vary during the measuremen...

Page 6: ...F The measured capacitance is indicated on the display 1 Set the selector switch to uF 2 Connect the test leads 3 After a short settling time the reading will be displayed automatically Zeroing of uF...

Page 7: ...over Remove the cells if the instrument is not going to be used for an extended period of time Fuse Checking and Replacement To check the instrument fuse switch to an insulation range and press the TE...

Page 8: ...1999mV a c 16 460 Hz 5 5 digits d c milliVolts zeroing Up to 9 9mV Input resistance 3M CAPACITANCE Measuring Range 0 1nF to 9 99uF Accuracy 3 2digits 0 2nF uF zeroing up to 10nF Basic and service erro...

Page 9: ...00 4 76 0 50 0 44 10 00 9 56 0 60 0 54 20 00 18 8 0 70 0 63 30 00 28 4 0 80 0 73 40 00 38 0 0 90 0 82 50 00 47 6 1 00 0 92 100 00 92 0 SAFETY The instruments meet the requirements for double insulatio...

Page 10: ...erating humidity 90 RH at 40 C max Storage temperature range 25 to 65 C Calibration Temperature 20 C Maximum altitude 2000m Dust and water protection IP54 Temperature coefficient 0 1 per C WEIGHT 742g...

Page 11: ...contact Megger Limited or Megger Archcliffe Road Valley Forge Corporate Center Dover 2621 Van Buren Avenue Kent CT17 9EN Norristown PA 19403 England U S A Tel 44 0 1304 502243 Tel 1 610 676 8500 Fax 4...

Page 12: ...l U S A Fax 1 214 331 7399 Z A Du Buisson de la Couldre 23 rue Eug ne Henaff 78190 TRAPPES Tel 33 0 1 30 16 08 90 FRANCE Fax 33 0 1 34 61 23 77 This instrument is manufactured in the United Kingdom Th...

Reviews: