Factory reset to Default settings
Factory default settings
SETUP - change test pass limits, test times and test lead resistance. SETUP is “test group based” as the PASS limit for a Class I insulation test is different to a Class II insulation test.
Factory Default Test Limits
Variant Model
R
pe
,
R
cont
(Ω)
R
pe
, R
cont
(Ω)
for Ext
lead
R
pe
, R
cont
(Ω)
for RCD
Class 1
R
iso
, R
ins
(MΩ)
Class 2
R
iso
, R
ins
(MΩ)
Ext lead
R
iso
, R
ins
(MΩ)
Class 1
I
ea
,
i
leak
(sub),
i
pe
, i
leak
(mA)
It, I
b
Class 2
i
ea
, It(sub)
(mA)
1xIΔN30
(ms)
5xIΔN30
(ms)
1xIΔN10
(ms)
5xIΔN10
(ms)
PAT120-UK
0.2
0.2
0.2
1
2
1
3.5
0.25
NA
NA
NA
NA
PAT150-UK
0.2
0.2
0.2
1
2
1
3.5
0.25
200
40
200
40
PAT120-DE, PAT120-CH,
PAT120-EU
0.3
0.3
0.3
1
2
1
3.5
0.5
NA
NA
NA
NA
PAT150-DE, PAT150-CH,
PAT150-EU
0.3
0.3
0.3
1
2
1
3.5
0.5
300
NA
300
NA
PAT150-AU
1
1
1
1
1
1
5
1
300
NA
40
NA