![Mega MEg40+ Manual Download Page 27](http://html1.mh-extra.com/html/mega/meg40/meg40_manual_1764028027.webp)
27
16MB memory card and memory card reader
2.5 Replacement of 16 MB memory cards in the measuring instrument MEg40
+
After the 16 MB memory card is taken out from the measuring instrument MEg40
+
, the
measured data recording is interrupted and this status is displayed alternately with the
measured data on the display MEg40
+
with the message „Memory module taken out“.
Then a new deleted memory card can be inserted, which will be marked by the
instrument (with the head of the running measurement) and the instrument shows the
message „Memory module inserted. Start of recording“ on the display for 1.5 sec. Then it
continues storing the measured data onto this card.
It is also possible to insert a not deleted card, which is marked by another instrument.
This wrong situation is alternately reported with the message „Memory module inserted.
Module not deleted“ and measured data are being lost until a deleted card is inserted.
Also repeated insertion of the memory card can be assumed, onto which measured data
were recorded before. In this case, the display shows the message „Memory module
inserted. Recording restored“ and the instrument continues recording the measured data
onto the repeatedly inserted memory card. The data measured during the time when the
card was not inserted will be lost.
3/ ManUFactUrEr
MEgA – Měřící Energetické Aparáty, a.s.
664 31 Česká 390, Czech Republic
tel: +420 545 214 988 • mail: [email protected] • web: http://www.e-mega.cz
Summary of Contents for MEg40+